Literature DB >> 30195416

Combination of Adaptive Statistical Iterative Reconstruction-V and Lower Tube Voltage During Craniocervical Computed Tomographic Angiography Yields Better Image Quality with a Reduced Radiation Dose.

Chih-Wei Chen1, Po-An Chen1, Chiung-Chen Chou1, Jui-Hsun Fu1, Po-Chin Wang1, Shuo-Hsiu Hsu1, Ping-Hong Lai2.   

Abstract

RATIONALE AND
OBJECTIVES: To evaluate image quality and radiation exposure when using the adaptive statistical iterative reconstruction-V (ASIR-V) algorithm for reconstructing craniocervical computed tomographic angiography images acquired at 100 kVp.
MATERIALS AND METHODS: We randomly divided 121 patients into three groups: group A (conventional protocol), 120 kVp with filtered back projection; group B, 120 kVp with 50% ASIR-V; and group C, 100 kVp with 50% ASIR-V. All patients underwent scans in a 256-slice computed tomography (CT) scanner. Radiation dose (volume CT dose index), dose-length product, and effective dose, objective parameters such as arterial attenuation value, signal-to-noise ratio, contrast-to-noise ratio, and noise obtained at head, neck, and shoulder levels were compared among the groups. Subjective image quality was independently assessed by two radiologists, and interobserver reliability was assessed using kappa analysis.
RESULTS: The radiation dose in group C was the lowest (p < 0.01) with a 40% reduction in volume CT dose index, dose-length product, and effective dose values compared to group A, and group C showed higher arterial attenuation than either group A or B (p < 0.01). Additionally, signal-to-noise ratio and contrast-to-noise ratio were higher and noise was lower in groups B and C than group A. Group C had better subjective image quality than groups A and B (p < 0.05), and the interobserver reliability between the two radiologists was high (k = 0.783).
CONCLUSION: Compared to the conventional protocol, using 50% ASIR-V and the 100 kVp protocol during craniocervical computed tomographic angiography yields better objective and subjective image quality at lower radiation doses.
Copyright © 2018 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Adaptive statistical iterative reconstruction-V; Craniocervical CTA; Image quality; Radiation dose

Year:  2018        PMID: 30195416     DOI: 10.1016/j.acra.2018.07.019

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  2 in total

1.  Low-dose CT angiography using ASiR-V for potential living renal donors: a prospective analysis of image quality and diagnostic accuracy.

Authors:  Woong Kyu Han; Joon Chae Na; Sung Yoon Park
Journal:  Eur Radiol       Date:  2019-08-30       Impact factor: 5.315

2.  Application of multiple injections of contrast agent in head and neck CT arteriovenous angiography in children: Study protocol clinical trial (SPIRIT Compliant).

Authors:  Shilong Tang; Guanping Zhang; Zhuo Chen; Xianfan Liu; Xiao Fan; Daisong Liu; Ling He
Journal:  Medicine (Baltimore)       Date:  2020-04       Impact factor: 1.817

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.