Influence of multiple reuse and resterilization cycles on the performance of a bipolar vessel sealing device (LigaSure) intended for single use.

Alexandra Kuvaldina1, Galina Hayes1, Julia Sumner1, Erica Behling-Kelly1.   

Abstract

OBJECTIVE: To determine the number of use/cleaning/resterilization cycles that can be safely applied to a vessel sealing device intended for single use (LigaSure). STUDY
DESIGN: Ex vivo study. SAMPLE POPULATION: LigaSure Small Jaw handsets (n = 6) and LigaSure Impact handsets (n = 6).
METHODS: Handsets underwent simulated splenectomy/cleaning/resterilization cycles until failure, defined as leaking vascular seal or blade retraction failure. Functional testing included assessment of vascular seal integrity, handset activation/tissue release, and cutting blade wear/retraction. Vascular seal failure was defined as a leak occurring at <300 mm Hg. Cycles to failure were recorded. Sealed vessels were evaluated by histology at first handset use and failure.
RESULTS: Vascular seals created with the Small Jaw handset failed at a mean (95% CI) of 17.2 cycles (9.6-24.8) and a minimum of 10 cycles. Vascular seals created with the Impact failed at a mean of 20 cycles (18.4-21.6) and a minimum of 17 cycles. The majority of seal failures (73%; 95% CI 39%-94%) immediate leaked during vessel filling. The rate of vascular seal failure increased after the initial failure. Failure was associated with histologic disparities in tissue apposition.
CONCLUSION: Repeated use and resterilization resulted in failure of the vascular seal due to inadequate tissue apposition after a minimum of 10 cycles. CLINICAL SIGNIFICANCE: Surgeons reusing and resterilizing LigaSure handsets (ForceTriad platform) should consider discarding handsets after 9 cycles for the Small Jaw and after 16 cycles for the Impact. Handsets should be immediately discarded after any intraoperative identification of vascular seal failure.
© 2018 The American College of Veterinary Surgeons.

Entities:  

Mesh:

Year:  2018        PMID: 30180278     DOI: 10.1111/vsu.12948

Source DB:  PubMed          Journal:  Vet Surg        ISSN: 0161-3499            Impact factor:   1.495


  3 in total

1.  Effect of multiple use and sterilization on sealing performance of bipolar vessel sealing devices.

Authors:  Silvia Gardeweg; Barbara Bockstahler; Gilles Duprè
Journal:  PLoS One       Date:  2019-08-20       Impact factor: 3.240

2.  LigaSure for the Creation of Bloodless Breast Pockets in Patients Undergoing Transaxillary Breast Augmentation.

Authors:  Min Kuk Kim; Sangdal Lee
Journal:  Plast Reconstr Surg Glob Open       Date:  2020-12-22

3.  Ex vivo comparison of sliding knot ligatures vs. haemostatic clips for equine small intestinal mesenteric vessel occlusion.

Authors:  Gessica Giusto; Marco Gandini
Journal:  BMC Vet Res       Date:  2020-08-12       Impact factor: 2.741

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.