ESTIMATION OF EFFECTIVE DOSE OF DENTAL X-RAY DEVICES.

Wang Qiang1, Fu Qiang1, Lin Lin1.   

Abstract

This study aims to estimate the effective doses of dental X-ray devices under common scanning protocols. After putting TLDs in the Alderson Radiation Therapy Phantom, we exposed the phantom under common scanning protocols of three dental X-ray devices, namely CBCT, dental panoramic machine and intraoral round cone device. Then effective doses were calculated using the measured absorbed doses of organs and tissues. Tissue weighting factors recommended by the ICRP were adopted in the calculation. Effective doses under common scanning protocols of three Dental X-ray devices were obtained. The effective dose of dental CT was 0.20 mSv, and that of dental panoramic machine and intraoral radiography were 0.013 and 0.0050 mSv, respectively. The tissue absorbed doses of dental CT scan were 0.63 mGy of brain, 7.7 mGy of salivary glands, 8.7 mGy of thyroid and 4.0 mGy of the lens of the eye. The tissue absorbed doses from dental panoramic machine are 0.62 mGy of salivary glands and 0.25 mGy of thyroid. And finally the tissue absorbed dose of intraoral radiography was 0.80 mGy of salivary gland. Among the three dental X-ray devices studied, dental CBCT scan can cause much higher effective dose than the other two. Brain, salivary glands, thyroid and the lens of the eye are tissues receiving relatively higher absorbed doses.
© The Author(s) 2018. Published by Oxford University Press. All rights reserved. For Permissions, please email: journals.permissions@oup.com.

Year:  2019        PMID: 30169836     DOI: 10.1093/rpd/ncy159

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  5 in total

1.  Ex-vivo imaging of buccal and oral periodontal bone with low-dose CBCT in porcine jaws.

Authors:  Maurice Ruetters; Holger Gehrig; Dorohtea Kronsteiner; Vanessa Weyer; Ti-Sun Kim; Christopher J Lux; Sinan Sen
Journal:  Dentomaxillofac Radiol       Date:  2021-07-08       Impact factor: 2.419

2.  A Review of Doses for Dental Imaging in 2010-2020 and Development of a Web Dose Calculator.

Authors:  Hawon Lee; Andreu Badal
Journal:  Radiol Res Pract       Date:  2021-12-10

3.  Imaging furcation defects with low-dose cone beam computed tomography.

Authors:  Maurice Ruetters; Holger Gehrig; Ti-Sun Kim; Valentin Bartha; Thomas Bruckner; Franz Sebastian Schwindling; Anna Felten; Christopher Lux; Sinan Sen
Journal:  Sci Rep       Date:  2022-04-26       Impact factor: 4.996

Review 4.  Indications for 3-D diagnostics and navigation in dental implantology with the focus on radiation exposure: a systematic review.

Authors:  Burkhard Kunzendorf; Hendrik Naujokat; Jörg Wiltfang
Journal:  Int J Implant Dent       Date:  2021-05-27

5.  Age-dependent decrease in dental pulp cavity volume as a feature for age assessment: a comparative in vitro study using 9.4-T UTE-MRI and CBCT 3D imaging.

Authors:  Maximilian Timme; Jens Borkert; Nina Nagelmann; Adam Streeter; André Karch; Andreas Schmeling
Journal:  Int J Legal Med       Date:  2021-04-26       Impact factor: 2.686

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.