Austin Adams1, Joely A Straseski2, Christopher M Lehman1,2, Lauren N Pearson1,2.
Abstract
BACKGROUND: Chemistry testing is requested for body fluid (BF) specimens despite the lack of assays approved by the US Food and Drug Administration (FDA). The criteria for categorizing fluids as transudate or exudate are not validated across analyzers.
OBJECTIVE: To compare BF chemical analysis and classification by different analyzers.
METHODS: We analyzed 10 pleural and 18 peritoneal fluids with corresponding plasma specimens using the Vitros 5,1 FS; Abbott ARCHITECT ci8200; and Roche Modular P platforms. Total protein (TP) and lactate dehydrogenase (LDH) were measured for pleural fluids. Light's criteria were applied. Albumin was measured for peritoneal specimens, and the plasma-ascites-albumin gradient was calculated.
RESULTS: TP results showed agreement. The Vitros LDH assay produced higher fluid:plasma ratios. Classification by Light's criteria resulted in 1 discrepancy (ARCHITECT). Albumin results showed agreement. There were 2 discrepant gradient interpretations (Vitros).
CONCLUSIONS: These data suggest that analyses of pleural and peritoneal fluids using these platforms are diagnostically interchangeable. © American Society for Clinical Pathology 2018. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.
BACKGROUND: Chemistry testing is requested for body fluid (BF) specimens despite the lack of assays approved by the US Food and Drug Administration (FDA). The criteria for categorizing fluids as transudate or exudate are not validated across analyzers.
OBJECTIVE: To compare BF chemical analysis and classification by different analyzers.
METHODS: We analyzed 10 pleural and 18 peritoneal fluids with corresponding plasma specimens using the Vitros 5,1 FS; Abbott ARCHITECT ci8200; and Roche Modular P platforms. Total protein (TP) and lactate dehydrogenase (LDH) were measured for pleural fluids. Light's criteria were applied. Albumin was measured for peritoneal specimens, and the plasma-ascites-albumin gradient was calculated.
RESULTS: TP results showed agreement. The Vitros LDH assay produced higher fluid:plasma ratios. Classification by Light's criteria resulted in 1 discrepancy (ARCHITECT). Albumin results showed agreement. There were 2 discrepant gradient interpretations (Vitros).
CONCLUSIONS: These data suggest that analyses of pleural and peritoneal fluids using these platforms are diagnostically interchangeable. © American Society for Clinical Pathology 2018. All rights reserved. For permissions, please e-mail: journals.permissions@oup.com.
Entities:
Keywords:
Light’s criteria; ascites-albumin gradient; body fluid; chemistry analytes; peritoneal; pleural
Mesh:
Substances:
Year: 2019
PMID: 30169773 DOI: 10.1093/labmed/lmy056
Source DB: PubMed Journal: Lab Med ISSN: 0007-5027