Characterization of PTW-31015 PinPoint ionization chambers in photon and proton beams.

A Carlino1, M Stock, N Zagler, M Marrale, J Osorio, S Vatnitsky, H Palmans.   

Abstract

The increased use of complex forms of radiotherapy using small-field photon and proton beams has invoked a growing interest in the use of micro-ionization chambers. In this study, 48 PTW-TM31015 PinPoint-type micro-ionization chambers that are used in the commissioning and patient specific QA of a proton pencil beam scanning (PBS) delivery system have been characterized in proton and high-energy photon beams. In both beam modalities, the entire set of PinPoint chambers was characterized by imaging them, by evaluating their stability using check source measurements, by experimentally determining the ion recombination, polarity effect and by cross calibrating them in terms of absorbed dose to water against Farmer-type ionization chambers. Beam quality correction factors were theoretically derived for both beam modalities. None of the chambers' check source readings drifted by more than 0.5% over a one year period. Beam quality correction factors for the 6 MV photon with reference to 60Co were on average 1.0  ±  0.5% lower than the theoretical values calculated according to the data and procedures outlined in IAEA TRS-398. While this difference is within the overall dosimetric uncertainty, it is significant considering only uncorrelated uncertainties indicating inconsistencies in the theoretical data. Beam quality correction factors for the 179.2 MeV proton beam with reference to 60Co were in good agreement with the theoretical data. Ion recombination and polarity correction factors were very consistent for all chambers with standard deviations of 0.2% or below show that findings from more comprehensive investigations in the literature can be considered as representative for all the chambers of this type. The characterization of 48 PinPoint-type micro-ionization chambers performed in this study provided a unique opportunity to investigate chamber-to-chamber variations of calibration, beam quality correction factors, ion recombination and polarity correction factors for an unprecedented sample size of chambers for both high-energy photon and proton beams.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 30152791     DOI: 10.1088/1361-6560/aadd39

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  3 in total

1.  Evaluation of electromagnetic and nuclear scattering models in GATE/Geant4 for proton therapy.

Authors:  Andreas F Resch; Alessio Elia; Hermann Fuchs; Antonio Carlino; Hugo Palmans; Markus Stock; Dietmar Georg; Loïc Grevillot
Journal:  Med Phys       Date:  2019-04-15       Impact factor: 4.071

2.  Computer-assisted beam modeling for particle therapy.

Authors:  Hermann Fuchs; Alessio Elia; Andreas F Resch; Peter Kuess; Armin Lühr; Marie Vidal; Loïc Grevillot; Dietmar Georg
Journal:  Med Phys       Date:  2020-12-25       Impact factor: 4.071

3.  Phantom design and dosimetric characterization for multiple simultaneous cell irradiations with active pencil beam scanning.

Authors:  Monika Clausen; Suphalak Khachonkham; Sylvia Gruber; Peter Kuess; Rolf Seemann; Barbara Knäusl; Elisabeth Mara; Hugo Palmans; Wolfgang Dörr; Dietmar Georg
Journal:  Radiat Environ Biophys       Date:  2019-09-20       Impact factor: 1.925

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.