| Literature DB >> 30152297 |
W J C Leong1, H W Yuen1.
Abstract
BACKGROUND: Dislocation of the cochlear implant magnet is an uncommon but recognised complication of magnetic resonance imaging in a patient with a cochlear implant.Case reportThis paper reports a case of cochlear implant magnet dislocation despite head bandaging. The patient subsequently underwent endoscopic repositioning of the magnet under general anaesthesia. The cochlear implant system was tested intra-operatively and confirmed to be functioning. The patient was well at follow up and the small wound healed well with no complications.Entities:
Keywords: Adult; Cochlear Implants; Endoscopy; Humans; Magnetic Resonance Imaging; Magnets; Postoperative Complications
Mesh:
Year: 2018 PMID: 30152297 DOI: 10.1017/S0022215118001421