Quantification of enhancement in contrast-enhanced spectral mammography using a custom-made quantifier tool (I-STRIP): A proof-of-concept study.

M B I Lobbes1, H K P Mulder2, M Rousch3, W H Backes2, J E Wildberger2, C R L P N Jeukens2.   

Abstract

PURPOSE: Contrast-enhanced spectral mammography (CESM) is diagnostically superior to full-field digital mammography. An important improvement for CESM would be the ability to quantify enhancement. In this proof-of-concept study we present a method for quantifying CESM enhancement. METHODS AND MATERIALS: We developed a custom-made quantifier tool (I-STRIP) containing five chambers, each filled with a different iodine mass thickness (IMT). CESM grey values of the recombined image (CGV) in the I-STRIP were used to quantify breast IMT. We evaluated the I-STRIP's accuracy using a dedicated breast phantom containing chambers with known IMT's. Furthermore, we tested the effect of the I-STRIP on image quality and clinical use in five patients. Retrospectively, we studied whether current CESM protocols could distinguish between malignant and benign lesions in terms of CGV.
RESULTS: Phantom experiments showed that quantification was independent of chamber height and size, phantom thickness and I-STRIP position for different IMT's (1.5, 3.0 and 7.5 mg l/cm2). Near the phantom's periphery accuracy was found to be lower due to the breast-within-breast artifact. In the clinical setting (n = 5), workflow and image quality were not influenced by the I-STRIP. The mean IMT of these invasive breast cancers was 2.1 mg l/cm2 (range 1.3-3.4 mg l/cm2). Malignant lesions showed significantly higher CGV's than benign lesions (p = 0.002).
CONCLUSION: We showed in both phantom and clinical experiments that CESM quantification is feasible, without influencing workflow or image quality. The current CESM imaging protocol seems to be able to distinguish between benign and malignant breast lesions in terms of CGV.
Copyright © 2018 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Breast cancer; Contrast agent; Contrast-enhanced spectral mammography; Mammography; Quantification

Mesh:

Substances:

Year:  2018        PMID: 30150032     DOI: 10.1016/j.ejrad.2018.07.021

Source DB:  PubMed          Journal:  Eur J Radiol        ISSN: 0720-048X            Impact factor:   3.528


  2 in total

1.  Quantitative analysis of contrast enhanced spectral mammography grey value for early prediction of pathological response of breast cancer to neoadjuvant chemotherapy.

Authors:  Dong Xing; Ning Mao; Jianjun Dong; Heng Ma; Qianqian Chen; Yongbin Lv
Journal:  Sci Rep       Date:  2021-03-15       Impact factor: 4.379

2.  Diagnostic value of the enhancement intensity and enhancement pattern of CESM to benign and malignant breast lesions.

Authors:  Xiaoxiao Chi; Lei Zhang; Dong Xing; Peiyou Gong; Qianqian Chen; Yongbin Lv
Journal:  Medicine (Baltimore)       Date:  2020-09-11       Impact factor: 1.817

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.