Rapid high-resolution T1 mapping using a highly accelerated radial steady-state free-precession technique.

Zhitao Li1,2, Ali Bilgin1,2,3, Kevin Johnson1, Jean-Philippe Galons1, Srinivasan Vedantham1, Diego R Martin1, Maria I Altbach1,3.   

Abstract

BACKGROUND: T1 mapping is often used in some clinical protocols. Existing techniques are limited in slice coverage, and/or spatial-temporal resolution, or require long acquisitions. Here we present a multi-slice inversion-recovery (IR) radial steady-state free precession (radSSFP) pulse sequence combined with a principal component (PC) based reconstruction that overcomes these limitations.
PURPOSE: To develop a fast technique for multi-slice high-resolution T1 mapping. STUDY TYPE: Technical efficacy study done prospectively. PHANTOM/
SUBJECTS: IR-radSSFP was tested in phantoms, five healthy volunteers, and four patients with abdominal lesions. FIELD STRENGTH/SEQUENCE: IR-radSSFP was implemented at 3T. ASSESSMENT: Computer simulations were performed to optimize the flip angle for T1 estimation; testing was done in phantoms using as reference an IR spin-echo pulse sequence. T1 mapping with IR-radSSFP was also assessed in vivo (brain and abdomen) and T1 values were compared with literature. T1 maps were also compared with a radial IR-FLASH technique. STATISTICAL TESTS: A two-tailed t-test was used to compare T1 values in phantoms. A repeatability study was carried out in vivo using Bland-Altman analysis.
RESULTS: Simulations and phantom experiments showed that a flip angle of 20˚ was optimal for T1 mapping. When comparing single to multi-slice experiments in phantoms there were no significant differences between the means T1 values (P = 0.0475). In vivo results show that T1 maps with spatial resolution as high as 0.69 mm × 0.69 mm × 2.00 mm (brain) and 0.83 mm × 0.83 mm × 3.00 mm (abdomen) can be generated for 84 brain slices in 3 min and 10 abdominal slices in a breath-hold; T1 values were comparable to those reported in literature. The coefficients of variation from the repeatability study were 1.7% for brain and 2.5-2.7% in the abdomen. DATA
CONCLUSION: A multi-slice IR-radSSFP technique combined with a PC-based reconstruction was demonstrated for higher resolution T1 mapping. This technique is fast, motion-insensitive and yields repeatable T1 values comparable to those in literature. LEVEL OF EVIDENCE: 2 Technical Efficacy: Stage 1 J. Magn. Reson. Imaging 2019;49:239-252.
© 2018 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  T1; T1 mapping; compressed sensing; model-based; radial MRI

Mesh:

Year:  2018        PMID: 30142230     DOI: 10.1002/jmri.26170

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  2 in total

1.  A multi-scale residual network for accelerated radial MR parameter mapping.

Authors:  Zhiyang Fu; Sagar Mandava; Mahesh B Keerthivasan; Zhitao Li; Kevin Johnson; Diego R Martin; Maria I Altbach; Ali Bilgin
Journal:  Magn Reson Imaging       Date:  2020-09-01       Impact factor: 2.546

2.  Rapid high-resolution volumetric T1 mapping using a highly accelerated stack-of-stars Look Locker technique.

Authors:  Zhitao Li; Zhiyang Fu; Mahesh Keerthivasan; Ali Bilgin; Kevin Johnson; Jean-Philippe Galons; Srinivasan Vedantham; Diego R Martin; Maria I Altbach
Journal:  Magn Reson Imaging       Date:  2021-03-17       Impact factor: 3.130

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.