Endotracheal tube cuff pressure changes during manual cuff pressure control manoeuvres: An in-vitro assessment.

Norbert Aeppli1, Bastian Lindauer1, Marc P Steurer2, Markus Weiss3, Alexander Dullenkopf1.   

Abstract

BACKGROUND: Endotracheal tube (ETT) cuffs are designed to seal the lower airway for precise ventilation and to protect against ingress of pathogens from the pharyngeal space. Therefore, a minimal continuous cuff pressure must be maintained. Aim of this study was to analyse the course of cuff pressure in an in-vitro model during manual cuff pressure control manoeuvres.
METHODS: An artificial trachea was intubated with an appropriately sized ETT and cuff pressure set to 20 cm H2 O. Thirty-two experienced ICU nurses each performed six cuff pressure control manoeuvres (three times in two different ETTs) using a manual cuff pressure manometer. Course of cuff pressure from connecting the manometer to disconnecting it from the cuff pilot balloon was recorded using a pressure transducer.
RESULTS: There were 190 cuff pressure control manoeuvres suitable for analysis. In all control manoeuvres a cuff pressure below 20 cm H2 O was noted. In 20.0% of the control manoeuvres the cuff pressure dropped below 10 cm H2 O. Cuff pressure drops were mainly caused by initially connecting the manometer to the pilot balloon, less frequently by manipulating the pressure gauge of the manometer. Disconnecting the manometer after the control manoeuvre caused a cuff pressure drop in 78.1% of cases, contributing to a final cuff pressure below 20 cm H2 O in 31.3% of control manoeuvres.
CONCLUSION: Routine manual cuff pressure control manoeuvres in ETT cuffs result in considerable cuff pressure drops. This may have an impact on silent aspiration of pharyngeal contents passing along the cuff into the lower airway.
© 2018 The Acta Anaesthesiologica Scandinavica Foundation. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  cuff pressure; endotracheal tube; monitoring

Year:  2018        PMID: 30132783     DOI: 10.1111/aas.13249

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  2 in total

1.  A Tapered Cuff Tracheal Tube Decreases the Need for Cuff Pressure Adjustment After Surgical Retraction During Anterior Cervical Spine Surgery: A Randomized Controlled, Double-Blind Trial.

Authors:  Yi-Shiuan Li; Elise Chia-Hui Tan; Yueh-Ju Tsai; Mercedes Susan Mandell; Shiang-Suo Huang; Ting-Yun Chiang; Wen-Cheng Huang; Wen-Kuei Chang; Ya-Chun Chu
Journal:  Front Med (Lausanne)       Date:  2022-06-29

2.  The Changes of Endotracheal Tube Cuff Pressure during Manual and Intermittent Controlling in Intensive Care Units.

Authors:  Roghieh Nazari; Christopher Boyle; Mojgan Panjoo; Mohammad Salehpour-Omran; Hamid Sharif Nia; Ameneh Yaghoobzadeh
Journal:  Iran J Nurs Midwifery Res       Date:  2019-12-27
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.