Literature DB >> 30132713

Histopathologic and physiologic effect of overlapping vs single coronary stents: impact of stent evolution.

Atsushi Sakamoto1, Sho Torii1, Hiroyuki Jinnouchi1, Renu Virmani1, Aloke V Finn1,2.   

Abstract

INTRODUCTION: Bare metal stents (BMS) and drug eluting stents (DES) have been deployed in single and overlap configurations, the latter mostly to cover long-lesions. Both type of stents in overlap configuration increases the risk of adverse events. AREAS COVERED: We present the rationale for either BMS or DES usage utilizing both preclinical and clinical studies for the use of single vs. overlapped stents. Further, employing experimental and pathological studies, we discuss the disadvantages of stent overlapping vs. single stenting and propose that vessel injury, local blood flow disturbance, higher drug/polymer dose are mechanisms leading to poor clinical outcomes including in-stent restenosis (ISR), delayed arterial healing, thrombosis, and hypersensitivity reactions. We also explore whether newer generation (2nd generation) DES with reduced strut thickness, optimized drug release profiles, better polymer biocompatibility minimize the disadvantages of stent overlap seen in 1st generation DES. Finally, we examine the reasons why fully bioabsorbable scaffolds resulted in increased adverse events. EXPERT COMMENTARY: DES innovations have minimized the risk of ISR and repeat revascularization. In contradistinction to 1st generation DES, major adverse cardiac events (MACE) is similar between single and overlapped stents in 2nd generation DES, suggestion stent design play an important role in clinical outcomes.

Entities:  

Keywords:  Bare metal stent; bioabsorbable scaffold; delayed healing; drug eluting stent; overlapping; percutaneous coronary intervention; restenosis; stent thrombosis

Mesh:

Year:  2018        PMID: 30132713     DOI: 10.1080/17434440.2018.1515012

Source DB:  PubMed          Journal:  Expert Rev Med Devices        ISSN: 1743-4440            Impact factor:   3.166


  1 in total

1.  Evaluation of Mechanical Performances of Stents with 38 mm Length in Long Lesion.

Authors:  Xiaoting Yue; Jiacheng Guo; Jianchao Zhang; Chang Cao; Zenglei Zhang; Deliang Shen; Junnan Tang; Jinying Zhang
Journal:  Biomed Res Int       Date:  2020-02-27       Impact factor: 3.411