Which plate results in better stability after segmental mandibular resection and fibula free flap reconstruction? Biomechanical analysis.

Si-Myung Park1, Jung-Woo Lee2, Gunwoo Noh3.   

Abstract

OBJECTIVE: This study investigated the biomechanical stability of 2 plate systems-mini-plates and reconstruction plates-in reconstruction with fibular free flaps. STUDY
DESIGN: The reconstruction models were constructed by using 2 types of plates in representative cases with segmental mandibular defect (C, L, LC1, LC2). In each model, a masticatory simulation approximating 3 clenching tasks was conducted, using the muscle forces adjusted to the mandible structure used in this study. In addition, to evaluate the sensitivity of the 2 plate systems for masticatory load changes, a sensitivity analysis was also performed by using finite element analysis.
RESULTS: The risks of plate fracture and screw loosening measured by stress concentrations were higher in the cases using mini-plates compared with those using reconstruction plates. Moreover, the mini-plate was more sensitive to varied loads compared with the reconstruction plate and was observed to have less flexibility to absorb external forces. Mini-plates also caused high strain values, indicating the risk of hypertrophy of bone around the screw holes.
CONCLUSIONS: The use of a reconstruction plate should result in more stable surgical outcomes in most cases, but we noted that the risk of atrophy may increase with the use of reconstruction plates because of lack of bone stimulation.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2018        PMID: 30126810     DOI: 10.1016/j.oooo.2018.05.048

Source DB:  PubMed          Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol


  4 in total

1.  Mechanical Fatigue Performance of Patient-Specific Polymer Plates in Oncologic Mandible Reconstruction.

Authors:  Julian Lommen; Lara Schorn; Christoph Sproll; Norbert R Kübler; Luis Fernando Nicolini; Ricarda Merfort; Ayimire Dilimulati; Frank Hildebrand; Majeed Rana; Johannes Greven
Journal:  J Clin Med       Date:  2022-06-09       Impact factor: 4.964

2.  Biomechanical Analysis of Various Reconstructive Methods for the Mandibular Body and Ramus Defect Using a Free Vascularized Fibula Flap.

Authors:  Xian Li; Chao Jiang; Hui Gao; Chunjuan Wang; Chao Wang; Ping Ji
Journal:  Biomed Res Int       Date:  2020-03-13       Impact factor: 3.411

3.  A small number of residual teeth after the mandibular resection of oral cancer is associated with titanium reconstruction plate exposure.

Authors:  Hiromi Hirohata; Toru Yanagawa; Shohei Takaoka; Kenji Yamagata; Kaoru Sasaki; Yoichiro Shibuya; Fumihiko Uchida; Satoshi Fukuzawa; Katsuhiko Tabuchi; Shogo Hasegawa; Naomi Ishibashi-Kanno; Mitsuru Sekido; Hiroki Bukawa
Journal:  Clin Exp Dent Res       Date:  2019-06-07

Review 4.  Complication of osteo reconstruction by utilizing free vascularized fibular bone graft.

Authors:  Qifeng Ou; Panfeng Wu; Zhengbing Zhou; Ding Pan; Ju-Yu Tang
Journal:  BMC Surg       Date:  2020-10-02       Impact factor: 2.102

  4 in total