Four-dimensional flow cardiovascular magnetic resonance in aortic dissection: Assessment in an ex vivo model and preliminary clinical experience.

Hector W de Beaufort1, Dipan J Shah2, Avni P Patel2, Matthew S Jackson2, Domenico Spinelli3, Eric Y Yang2, Mohamad G Ghosn2, Kyle Autry2, Stephen R Igo2, Alan B Lumsden2, Stephen H Little2, Santi Trimarchi4, Jean Bismuth5.   

Abstract

OBJECTIVE: Four-dimensional flow cardiovascular magnetic resonance may improve assessment of hemodynamics in patients with aortic dissection. The purpose of this study was to evaluate the feasibility and accuracy of 4-dimensional flow cardiovascular magnetic resonance assessment of true and false lumens flow.
METHODS: Thirteen ex vivo porcine aortic dissection models were mounted to a flow loop. Four-dimensional flow cardiovascular magnetic resonance and 2-dimensional phase-contrast cardiovascular magnetic resonance measurements were performed, assessed for intraobserver and interobserver variability, and compared with a reference standard of sonotransducer flow volume measurements. Intraobserver and interobserver variability of 4-dimensional flow cardiovascular magnetic resonance were also assessed in 14 patients with aortic dissection and compared with 2-dimensional phase-contrast cardiovascular magnetic resonance.
RESULTS: In the ex vivo model, the intraobserver and interobserver measurements had Lin's correlation coefficients of 0.98 and 0.96 and mean differences of 0.17 (±3.65) mL/beat and -0.59 (±5.33) mL/beat, respectively; 4-dimensional and sonotransducer measurements had a Lin's concordance correlation coefficient of 0.95 with a mean difference of 0.35 (±4.92) mL/beat, respectively. In patients with aortic dissection, the intraobserver and interobserver measurements had Lin's concordance correlation coefficients of 0.98 and 0.97 and mean differences of -0.95 (±8.24) mL/beat and 0.62 (±10.05) mL/beat, respectively; 4-dimensional and 2-dimensional flow had a Lin's concordance correlation coefficient of 0.91 with a mean difference of -9.27 (±17.79) mL/beat because of consistently higher flow measured with 4-dimensional flow cardiovascular magnetic resonance in the ascending aorta.
CONCLUSIONS: Four-dimensional flow cardiovascular magnetic resonance is feasible in patients with aortic dissection and can reliably assess flow in the true and false lumens of the aorta. This promotes potential future work on functional assessment of aortic dissection hemodynamics.
Copyright © 2018 The American Association for Thoracic Surgery. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  4D-flow MRI; aorta; aortic dissection; ex vivo model

Mesh:

Year:  2018        PMID: 30121136     DOI: 10.1016/j.jtcvs.2018.06.022

Source DB:  PubMed          Journal:  J Thorac Cardiovasc Surg        ISSN: 0022-5223            Impact factor:   5.209


  5 in total

1.  Parametric Hemodynamic 4D Flow MRI Maps for the Characterization of Chronic Thoracic Descending Aortic Dissection.

Authors:  Kelly Jarvis; Judith T Pruijssen; Andre Y Son; Bradley D Allen; Gilles Soulat; Alireza Vali; Alex J Barker; Andrew W Hoel; Mark K Eskandari; S Chris Malaisrie; James C Carr; Jeremy D Collins; Michael Markl
Journal:  J Magn Reson Imaging       Date:  2019-11-12       Impact factor: 4.813

2.  Detection and Hemodynamic Evaluation of Flap Fenestrations in Type B Aortic Dissection with 4D Flow MRI: Comparison with Conventional MRI and CTA.

Authors:  Bradley D Allen; Pascal J Aouad; Nicholas S Burris; Amir Ali Rahsepar; Kelly B Jarvis; Christopher J Francois; Alex J Barker; S Chris Malaisrie; James C Carr; Jeremy D Collins; Michael Markl
Journal:  Radiol Cardiothorac Imaging       Date:  2019-04-25

3.  False lumen pressure estimation in type B aortic dissection using 4D flow cardiovascular magnetic resonance: comparisons with aortic growth.

Authors:  David Marlevi; Julio A Sotelo; Ross Grogan-Kaylor; Yunus Ahmed; Sergio Uribe; Himanshu J Patel; Elazer R Edelman; David A Nordsletten; Nicholas S Burris
Journal:  J Cardiovasc Magn Reson       Date:  2021-05-13       Impact factor: 5.364

4.  Color-coded summation images for the evaluation of blood flow in endovascular aortic dissection fenestration.

Authors:  Anne Marie Augustin; Franziska Wolfschmidt; Thilo Elsässer; Alexander Sauer; Alexander Dierks; Thorsten Alexander Bley; Ralph Kickuth
Journal:  BMC Med Imaging       Date:  2022-02-04       Impact factor: 1.930

5.  The impact of 4D-Flow MRI spatial resolution on patient-specific CFD simulations of the thoracic aorta.

Authors:  Zinedine Khatir; Amirul Khan; Malenka Bissell; Molly Cherry
Journal:  Sci Rep       Date:  2022-09-06       Impact factor: 4.996

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.