Design and performance of the micro-dose calibrator.

Stephen Adler1, Peter Choyke.   

Abstract

A new class of dose calibrator called a micro-dose calibrator, designed to measure radioactivity rates in the range between 1 kBq up to 370 kBq, has been designed and an 8 NaI(Tl)/photo multiplier tube (PMT) module prototype built to evaluate its performance. The design of the micro-dose calibrator is based on the concept of a segmented well counter. By segmenting a single crystal well counter into multiple crystals or segments, with each segment operated and read out independently, one can extend the operational range of a standard single crystal well counter while maintaining its accuracy and spectrographic capabilities. The micro-dose calibrator is operated in N-fold coincidence mode to ensure that a single radioactive decay resulting in multiple segments detecting gamma ray events are counted as a single decay. The N fold coincidence detection is achieved using a pipeline coincidence processing algorithm implemented in software as part of a multi-threaded acquisition system. A tri-functional model pulse pileup dead time correction is applied to extend the operational range of the micro-dose calibrator. The micro-dose calibrator achieved an operation range between 1 kBq up to 3.4 MBq for 18F with a 1% linearity precision using a 60 s integration time window. Over a 9 d period, a daily count rate measure of a 137Cs point source had a coefficient of variation of 0.17% with only one high voltage calibration performed on day 0. The central 20 mm up to 35 mm axial length in the sample well showed less than 1% change in count rate depending on the radionuclide. Using a 0.5 ml insulin syringe, a volume of dose between 0.13 ml up to 0.50 ml showed less than 1% deviation from volumetric linearity. A 20 cc scintillation vial showed similar volumetric linearity performance for radionuclide solution volumes between 4 cc up to 20 cc.

Entities:  

Mesh:

Substances:

Year:  2018        PMID: 30118440      PMCID: PMC7984724          DOI: 10.1088/1361-6560/aadb3b

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  An improved gamma well counter for radioactive tracer applications.

Authors:  F E ten Haaf; M L Verheijke
Journal:  Int J Appl Radiat Isot       Date:  1976-02

Review 2.  Routine quality control of clinical nuclear medicine instrumentation: a brief review.

Authors:  Pat Zanzonico
Journal:  J Nucl Med       Date:  2008-07       Impact factor: 10.057

  2 in total
  3 in total

1.  Cell radiolabeling with acoustophoresis cell washing.

Authors:  Stephen S Adler; Emmanuel C Nyong; Raisa A Glabman; Peter L Choyke; Noriko Sato
Journal:  Sci Rep       Date:  2022-06-01       Impact factor: 4.996

2.  Sodium Fluoride-18 and Radium-223 Dichloride Uptake Colocalize in Osteoblastic Mouse Xenograft Tumors.

Authors:  Tim E Phelps; Jyoti Roy; Michael V Green; Jurgen Seidel; Kwamena E Baidoo; Stephen Adler; Elijah F Edmondson; Donna Butcher; Jennifer L Matta; Anita T Ton; Karen Wong; Shan Huang; Ling Ren; Amy K LeBlanc; Peter L Choyke; Elaine M Jagoda
Journal:  Cancer Biother Radiopharm       Date:  2021-02-25       Impact factor: 3.099

3.  A Study of 219Rn Outgassing and 211Pb Contamination from 223Ra In dry, Liquid, and Murine Tissue Samples.

Authors:  Stephen Adler; Kwamena Baidoo; Elaine Jagoda; Tim Phelps; Jyoti Roy; Jurgen Seidel; Frank Lin; Peter Choyke
Journal:  Health Phys       Date:  2020-02       Impact factor: 2.922

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.