| Literature DB >> 30114501 |
Heepeel Chang1, Neal C Hadro2, Marc A Norris2, Sang Won Rhee2, Marvin E Morris2.
Abstract
Despite recent publications indicating billowing (aneurysmal degeneration of the outer fabric material manifesting as a bulging sac of contained contrast due to weakening of the fabric material between the metal struts) as a benign finding, we present a unique case of billowing that resulted in aneurysmal rupture with the AFX2® stent graft (Endologix, Inc., Irvine, CA). AFX2® consists of an inner metallic endoskeleton with multiple metallic struts covered by a polytetrafluoroethylene graft fabric. It allows the flow of blood or contrast beyond the border of the metallic endoskeleton without evidence of endoleak or rupture secondary to the outer cover being anchored to the endoskeleton with polypropylene sutures only at the proximal and distal ends. Our report raises the concern that although billowing may be a relatively common manifestation with Endologix stent grafts, clinicians have to be diligent as the phenomenon may progress to abdominal aortic aneurysm rupture.Entities:
Mesh:
Year: 2018 PMID: 30114501 DOI: 10.1016/j.avsg.2018.06.019
Source DB: PubMed Journal: Ann Vasc Surg ISSN: 0890-5096 Impact factor: 1.466