Effect of Cell Inner Pressure on Deposition Volume in Microinjection.

Xian Wang1,2, Qili Zhao1, Li Wang1, Jun Liu1, Huayan Pu3, Shaorong Xie3, Changhai Ru4, Yu Sun1,2,5.   

Abstract

Microinjection is a widely used technique for introducing exogenous materials into cells. Many applications of microinjection, such as gene editing and drug testing, rely on the accurate control of the deposition volume. However, the deposition volume in microinjection is presently calibrated in an open medium without considering the cell inner pressure effect, which we experimentally show in this paper that it can induce an error as large as 30% between the actual deposition volume and the set volume. In this work, the relationship between the cell inner pressure and the deposition volume was analytically modeled and experimentally validated. On the basis of the developed model, the cell inner pressure of a given cell type can be well estimated from the injection pressure and the resulting deposition volume. The quantitated cell inner pressure is then used to reduce the error between the set volume and the actual deposition volume. Experiments conducted on human bladder cancer cells (T24 and RT4) showed that T24 cells have a higher inner pressure than RT4 cells (405 ± 45 Pa vs 341 ± 34 Pa), and after compensating for the cell inner pressure, the error between the intended set volume and the actual deposition volume into a cell became less than 3%.

Entities:  

Year:  2018        PMID: 30095920     DOI: 10.1021/acs.langmuir.8b02102

Source DB:  PubMed          Journal:  Langmuir        ISSN: 0743-7463            Impact factor:   3.882


  1 in total

Review 1.  Recent advances in critical nodes of embryo engineering technology.

Authors:  Youwen Ma; Mingwei Gu; Liguo Chen; Hao Shen; Yifan Pan; Yan Pang; Sheng Miao; Ruiqing Tong; Haibo Huang; Yichen Zhu; Lining Sun
Journal:  Theranostics       Date:  2021-05-25       Impact factor: 11.556

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.