Hiroto Maeda1, Kaori Hayashi2, Takayuki Ishige3, Tomohiro Sunagawa1, Shigeko Tanigawa1, Mika Mishina4, Toshifumi Watanabe4, Kazuyuki Sogawa2.
Abstract
With the increasing number of cats kept as pets, opportunities to treat cats with lower urinary tract disease (LUTD) have recently increased in the clinical veterinary field. Urine samples collected from 50 cats with bacterial cystitis brought to Maeda Veterinary Hospital between August 10, 2015 and March 31, 2017 were used in the study. Sample preparation of the urine was performed using a MALDI Sepsityper kit and rapid BACpro. To identify the isolates, MALDI-TOF MS was performed on an AutoFlex TOF/TOF mass spectrometer. MALDI-TOF MS using rapid BACpro for pretreatment was found to be a quick and reliable method for identification of bacteria from infected urine, with a shortened analysis time enabling earlier and more accurate selection of antibiotics for treatment of feline LUTD.Entities:
Keywords: cat; lower urinary tract disease; matrix-assisted laser desorption ionization-time of flight mass spectrometry; urine
Mesh:
Year: 2018 PMID: 30089741 PMCID: PMC6207508 DOI: 10.1292/jvms.18-0145
Source DB: PubMed Journal: J Vet Med Sci ISSN: 0916-7250 Impact factor: 1.267
MALDI-TOF MS scores for different Escherichia coli and Enterococcus faecalis bacteria counts
| Kit | Microorganism | MALDI-TOF MS score for bacteria count
(CFU/m | ||||||
|---|---|---|---|---|---|---|---|---|
| 1.0 × 107 | 5.0 × 106 | 1.0 × 106 | 5.0 × 105 | 1.0 × 105 | 5.0 × 104 | 1.0 × 104 | ||
| MALDI Sepsityper | 2.321 | 2.301 | 2.284 | 2.184 | 2.105 | 1.988 | 1.904 | |
| 2.122 | 2.115 | 1.986 | 1.862 | 1.792 | 1.662 | 1.624 | ||
| Rapid BACpro | 2.398 | 2.321 | 2.299 | 2.186 | 2.134 | 1.991 | 1.972 | |
| 2.182 | 2.182 | 2.162 | 2.084 | 2.051 | 1.742 | 1.722 | ||
Urine culture results of 39 specimens analyzed by MALDI-TOF MS
| Urine culture (No. of isolates) | MALDI sepsityper (No. of isolates) | Rapid BACpro (No. of isolates) |
|---|---|---|
| Growth of colonies (N=37) | ||
| 1-colony morphology (N=29) | Positive with same identification (25)a) | Positive with same identification (25)a) |
| Not reliable identification (4)b) | Not reliable identification (4)b) | |
| 2-colony morphology (N=8) | Not reliable identification (8) | |
| Not reliable identification (1)c) | ||
| No growth (N=13) |
a) A total of 25 correct identifications at the species level. b) Four cases were cultures with <1.0 × 105 CFU/ml. c) Cases of E. faecalis and E. cloacae were cultures with <1.0 × 105 CFU/ml.
Identification by MALDI-TOF MS of 16S rRNA and conventional identification of 45 microorganisms causing cystitis with bacterial counts ≥1.0 × 105 CFU/ml
| Conventional method | MALDI-TOF MS method | 16S rRNA method | |
|---|---|---|---|
| MALDI sepsityper kit | Rapid BACpro | ||
| No reliable identification (2) | |||
| No reliable identification (1) | No reliable identification (1) | ||
| Total gram-negative bacteria (31) | 90.3% (28/31) | 96.7% (30/31) | 100.0% (31/31) |
| No reliable identification (9) | |||
| No reliable identification (1) | |||
| No reliable identification (2) | No reliable identification (2) | ||
| Total gram-positive cocci (14) | 21.4% (3/14) | 78.5% (11/14) | 100.0% (14/14) |
| Total (45) | 68.8% (31/45) | 91.1% (41/45) | 100.0% (45/45) |