| Literature DB >> 30089645 |
Danka Pavliakova1, Peter C Giardina2, Soraya Moghazeh2, Shite Sebastian3, Maya Koster2, Viliam Pavliak2, Andrew McKeen4, Roger French4, Kathrin U Jansen2, Michael Pride2.
Abstract
A Luminex-based direct immunoassay (dLIA) platform has been developed to replace the standardized pneumococcal enzyme-linked immunosorbent assay platform. The multiplex dLIA simultaneously measures the concentration of serum immunoglobulin G (IgG) antibodies specific for pneumococcal capsular polysaccharide (PnPS) serotypes 1, 3, 4, 5, 6A, 6B, 7F, 9V, 14, 18C, 19A, 19F, and 23F. The assay uses poly-l-lysine (PLL)-conjugated PnPS, chemically coupled to spectrally distinct Luminex microspheres. Assay validation experiments were performed using residual human serum samples obtained from 13-valent pneumococcal conjugate vaccine (13vPnC) clinical studies. Assay results are expressed as IgG antibody concentrations in micrograms per milliliter using the international reference serum, 007sp. The lower limit of quantitation (LLOQ) for all serotypes covered in the 13-plex dLIA fell within the range of 0.002 to 0.038 µg/ml serum IgG. The difference between the lower limit and upper limit of the assay range was >500-fold for all serotypes, and assay variability was <20% relative standard deviation (RSD) for all serotypes. IgG antibody measurements were shown to be serotype-specific (some cross-reactivity was observed only between the structurally related serotypes 6A and 6B as well as 19A and 19F), and no interference was observed between the serotypes when the assay was performed in the 13-plex format compared to the singleplex assays. The 13-plex dLIA platform developed by Pfizer Inc. generates up to 143 test results in a single 96-well plate and is a suitable replacement of the enzyme-linked immunosorbent assay (ELISA) platform for evaluating vaccine clinical trials.IMPORTANCE The pneumococcal enzyme-linked immunosorbent assay (ELISA) measures IgG antibodies in human serum, and it is an important assay that supports licensure of pneumococcal vaccines. The immune correlate of protection, 0.35 µg/ml of IgG antibodies, was determined by the ELISA method. Pfizer has developed a new Luminex-based assay platform to replace the ELISA. These papers describe the important work of (i) validating the Luminex-based assay and (ii) bridging the immune correlate of protection (0.35 µg/ml IgG) to equivalent values reported by the Luminex platform.Entities:
Keywords: Luminex; development; multiplex; pneumococcal; validation
Mesh:
Substances:
Year: 2018 PMID: 30089645 PMCID: PMC6083093 DOI: 10.1128/mSphere.00128-18
Source DB: PubMed Journal: mSphere ISSN: 2379-5042 Impact factor: 4.389
Results from assay robustness DoEs
| Serotype | RSD (%) | ||||||
|---|---|---|---|---|---|---|---|
| Day | Primary Ab | Primary Ab | Secondary Ab | Secondary Ab | Residual | Total | |
| 1 | 0.00 | 0.00 | 0.00 | 0.00 | 2.80 | 13.25 | 13.95 |
| 3 | 1.51 | 5.52 | 12.97 | 0.00 | 1.61 | 10.87 | 19.07 |
| 4 | 0.62 | 4.40 | 0.00 | 0.00 | 3.07 | 14.17 | 15.72 |
| 5 | 0.00 | 6.42 | 0.00 | 1.16 | 1.54 | 14.01 | 16.32 |
| 6A | 0.52 | 6.73 | 11.69 | 0.00 | 2.28 | 13.87 | 21.01 |
| 6B | 0.00 | 0.00 | 3.29 | 0.94 | 2.06 | 12.13 | 13.35 |
| 7F | 0.61 | 9.29 | 8.20 | 1.13 | 0.86 | 13.61 | 19.97 |
| 9V | 1.95 | 1.51 | 0.00 | 0.29 | 1.79 | 11.31 | 12.28 |
| 14 | 0.00 | 0.00 | 0.00 | 0.00 | 3.02 | 12.79 | 13.65 |
| 18C | 0.50 | 0.61 | 0.00 | 0.00 | 1.83 | 10.75 | 11.61 |
| 19A | 0.95 | 0.04 | 3.97 | 0.00 | 1.34 | 13.55 | 18.89 |
| 19F | 1.29 | 3.96 | 0.00 | 0.00 | 2.63 | 12.42 | 13.80 |
| 23F | 0.53 | 0.00 | 6.31 | 0.00 | 2.89 | 11.59 | 13.86 |
Primary and secondary antibody (Ab) incubation times, 60, 90, and 120 min; primary and secondary Ab incubation temperatures, 18, 20, and 25°C. Factorial design of experiments (DoEs) was performed to evaluate assay robustness. The following factors were studied: day (analyst), primary antibody incubation time and temperature, and secondary antibody incubation time and temperature. Primary and secondary antibody incubation times were evaluated in a range from 60 to 120 minutes and primary and secondary incubation temperature in a range from 18 to 25oC. Percent relative standard deviation (%RSD) was estimated for variations in test sera concentrations due to the factors evaluated. For each serotype, a variance component analysis was carried out to estimate the variability of each factor. The individual components of %RSD are not additive, because the effects of each component have a different weight on the total %RSD. The %RSD total is calculated by using the following formula (36): RSD = log(1 + %RSD2).
FIG 1 Dynamic ranges of anti-PnPS IgG Reference standard serum 007sp for each of 13 serotypes. MFI signal data for the reference standard curve are on the y axis, and specific IgG antibody concentration data are on the x axis. The blue line indicates the assay range lower limit as determined in the assay validation; the red line indicates the assay range upper limit.
Specificity of 13-plex dLIA
| Pneumococcal | % competition of reference standard serum 007sp with PnPs as | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | 3 | 4 | 5 | 6A | 6B | 7F | 9V | 14 | 18C | 19A | 19F | 23F | |
| 1 | −1 | 1 | 1 | 1 | 1 | 6 | 8 | 4 | −3 | −5 | −4 | 1 | |
| 3 | −4 | −1 | −2 | −4 | −5 | −1 | 2 | 3 | −7 | −1 | −2 | 0 | |
| 4 | 1 | −1 | 1 | −3 | 12 | 6 | 12 | 3 | −3 | −4 | 11 | 17 | |
| 5 | 8 | −2 | 1 | 1 | 1 | 8 | 6 | 5 | −4 | −5 | −5 | −3 | |
| 6A | −3 | 0 | 4 | 3 | 88 | 6 | 2 | 3 | −3 | −2 | 0 | 4 | |
| 6B | −4 | −1 | 2 | 0 | 38 | 6 | 3 | 3 | −3 | −3 | −3 | 7 | |
| 7F | −4 | 0 | 1 | −1 | 0 | 1 | 3 | 4 | 0 | −5 | −3 | −3 | |
| 9V | 2 | −2 | −2 | −1 | −1 | 0 | 6 | 1 | −3 | 3 | −3 | 0 | |
| 14 | −3 | 0 | 2 | 0 | 1 | 1 | 7 | 5 | −2 | −3 | −2 | 0 | |
| 18C | −2 | −1 | 0 | 2 | −3 | 3 | 3 | 14 | 3 | 1 | 0 | 2 | |
| 19A | −2 | 0 | 1 | 1 | 1 | 2 | 7 | 9 | 5 | 1 | 26 | −2 | |
| 19F | 14 | −1 | 3 | 1 | −1 | −1 | 5 | 15 | 5 | −4 | 33 | 0 | |
| 23F | −3 | 0 | 1 | 1 | 0 | 4 | 5 | 1 | 4 | 6 | −1 | 0 | |
Values shown represent percent reduction of IgG antibody concentration in 007sp reference standard serum in the presence of PnPS competitors at a concentration of 3 µg/ml. Bold data represent PnPs serotypes with homologous inhibition.
Comparison of mean IgG results between 13-plex and singleplex dLIA
| Pneumococcal | % bias of mean IgG concn (13-plex over 1-plex) | |||
|---|---|---|---|---|
| %RSD | GMR | Lower 90% confidence | Upper 90% confidence | |
| 1 | 11.0 | 97.7 | 94.4 | 101.0 |
| 3 | 28.3 | 103.9 | 95.8 | 112.8 |
| 4 | 21.9 | 94.7 | 88.7 | 101.0 |
| 5 | 14.6 | 96.2 | 92.0 | 100.5 |
| 6A | 12.1 | 95.5 | 92.2 | 98.8 |
| 6B | 14.3 | 98.9 | 95.0 | 103.0 |
| 7F | 12.6 | 94.6 | 91.2 | 98.2 |
| 9V | 23.8 | 96.4 | 89.9 | 103.5 |
| 14 | 10.4 | 94.8 | 91.2 | 98.4 |
| 18C | 13.0 | 91.1 | 87.6 | 94.7 |
| 19A | 12.1 | 90.1 | 86.5 | 93.8 |
| 19F | 12.2 | 99.3 | 95.4 | 103.4 |
| 23F | 15.2 | 101.5 | 97.1 | 106.0 |
Data represent results of two one-sided 5% interval tests.
Assay range based on accuracy, dilutional linearity, and precision
| Serotype | Well concn | Assay range (ng/ml) | ||||||
|---|---|---|---|---|---|---|---|---|
| Accuracy | Dilutional linearity | Precision | Lower limit | Upper limit | ||||
| Lower limit | Upper limit | Lower limit | Upper limit | Lower limit | Upper limit | |||
| 1 | 0.010 | 21.250 | 24.360 | 0.005 | ||||
| 3 | 0.004 | 10.730 | 0.002 | 9.462 | ||||
| 4 | 0.012 | 33.300 | 0.007 | 33.208 | ||||
| 5 | 0.013 | 26.000 | 0.006 | 19.978 | ||||
| 6A | 0.032 | 78.600 | 77.371 | 0.018 | ||||
| 6B | 0.011 | 47.070 | 0.007 | 46.888 | ||||
| 7F | 0.009 | 41.500 | 0.006 | 34.277 | ||||
| 9V | 0.010 | 32.200 | 0.007 | 30.040 | ||||
| 14 | 0.025 | 56.621 | 0.011 | 54.927 | ||||
| 18C | 0.006 | 18.250 | 0.003 | 18.868 | ||||
| 19A | 0.017 | 34.675 | 40.421 | 0.008 | ||||
| 19F | 0.008 | 36.525 | 0.004 | 36.322 | ||||
| 23F | 0.013 | 30.740 | 0.008 | 37.795 | ||||
The lower assay range limit was determined by the most conservative value (highest value for lower limit and lowest value for upper limit) of accuracy, dilution linearity, and precision, highlighted in bold. All of the values represent non-dilution-adjusted “well concentrations.”
Data represent non-dilution-adjusted concentrations.
Final assay range and lower limit of quantitation (dilution adjusted)
| Serotype | Dilution adjusted assay range (ng/ml) | Fold difference | ||
|---|---|---|---|---|
| Lower limit (ARLL) | Upper limit (ARUL) | LLOQ (ng/ml) | ||
| 1 | 0.8 | 947.5 | 2 | 1,184 |
| 3 | 0.4 | 362.5 | 4 | 906 |
| 4 | 1.0 | 1,457.5 | 5 | 1,458 |
| 5 | 1.1 | 938.8 | 2 | 853 |
| 6A | 1.7 | 3,410.4 | 5 | 2,006 |
| 6B | 0.9 | 2,262.5 | 15 | 2,514 |
| 7F | 0.5 | 866.1 | 3 | 1,732 |
| 9V | 0.8 | 827.1 | 13 | 1,034 |
| 14 | 3.6 | 2,374.4 | 5 | 660 |
| 18C | 0.6 | 831.6 | 2 | 1,386 |
| 19A | 1.4 | 1,733.8 | 38 | 1,238 |
| 19F | 0.7 | 1,735.1 | 12 | 2,479 |
| 23F | 1.4 | 1,487.5 | 9 | 1,063 |
The assay range determined by accuracy, dilution linearity, and precision was adjusted by a sample dilution factor of 50. LLOQ was determined from the precision results from the lowest (1:50) sample dilutions. This approach ensures that the LLOQ reflects the ability of the assay to precisely measure samples in a high-serum matrix.
Intermediate precision
| Serotype | Analyst | Day | Bead lot | Residual | Total | |
|---|---|---|---|---|---|---|
| 1 | 702 | 4.2 | 4.6 | 0.4 | 9.8 | 11.6 |
| 3 | 702 | 0.0 | 4.2 | 1.8 | 7.5 | 8.8 |
| 4 | 704 | 2.4 | 4.1 | 1.7 | 9.1 | 10.4 |
| 5 | 702 | 2.9 | 4.7 | 0.3 | 8.9 | 10.5 |
| 6A | 704 | 2.1 | 3.6 | 0.0 | 9.0 | 9.9 |
| 6B | 704 | 2.9 | 4.5 | 1.2 | 9.0 | 10.6 |
| 7F | 702 | 2.7 | 4.5 | 1.7 | 7.8 | 9.6 |
| 9V | 704 | 2.7 | 3.5 | 0.0 | 9.9 | 10.9 |
| 14 | 703 | 4.1 | 4.8 | 2.2 | 8.6 | 10.9 |
| 18C | 687 | 2.6 | 4.8 | 0.9 | 9.3 | 10.8 |
| 19A | 686 | 5.0 | 3.7 | 12.2 | 13.1 | 19.0 |
| 19F | 687 | 2.5 | 4.6 | 0.0 | 10.4 | 11.6 |
| 23F | 684 | 1.5 | 5.3 | 2.6 | 9.6 | 11.4 |
n, number of observations, including replicates.
FIG 2 Microsphere stability. Data represent IgG GMC of 44-member serum panel tested on the microspheres over the 40-week period. A proficiency panel of 44 human sera with IgG concentrations ranging from low to high was evaluated on the same microsphere lot over the 40-week period. A single-serum panel was prepared for all 13 serotypes. Each panel (representing one serotype) shows the reported IgG concentrations (y axis) of the proficiency panel IgG GMC, as a red dot, over time (x axis). A linear trend line is applied to each chart.
FIG 3 Proficiency panel performance in 13-plex dLIA. A proficiency panel of 44 sera with IgG concentrations ranging from low to high was evaluated post-assay validation quarterly for 1 year. Each panel (representing one serotype) shows the reported IgG concentration (y axis) of each of member of proficiency panel sera over time (x axis) as a black triangle. GMCs of the serum panels and standard deviation limits are indicated by solid red lines. Assay results are indicated in chronological order from left to right (first quarter 2015 to fourth quarter 2015).
Summary of microsphere coupling conditions
| Parameter | Value |
|---|---|
| EDC/sNHS concn (µg/ml) | 5,000 |
| Microsphere activating time (min) | 20 |
| Coupling buffer | 50 mM HEPES |
| PnPS-PLL coupling concn (µg/ml) | 0.5–1 µg/ml (based on serotype) |
| Coupling incubation time (min) | 120 |
| Bead concn in coupling reaction | 1.25 × 107/0.5 ml |
| Storage buffer | 1% BSA–DPBS |