An automated micro-separation system for the chromatographic removal of uranium matrix for trace element analysis by ICP-OES.

E Miller Wylie1, Benjamin T Manard2, C Derrick Quarles3, Lisa A Meyers2, Ning Xu2.   

Abstract

An automated, miniaturized, off-line separation technique is presented here using an Elemental Scientific Inc. microFAST MC system with UTEVA resin to extract the uranium matrix from its trace element impurities in aqueous media. The collected fractions were analyzed for ~ 30 trace elements using inductively coupled plasma - optical emission spectroscopy. Ten replicate samples were processed with a single column resulting in precision ranging from 3.3% to 6.2% relative standard deviation with regards to the trace element recoveries. Accuracy, with respect to trace element concentrations in the U3O8 Certified Reference Material 124-1, resulted in an average of 13.9% relative deviation while accuracy to the Canadian U3O8 reference material, CUP-2, resulted in an average relative deviation of 8.6%. The total separation time of this automated process was reduced to ~ 30 min per sample while employing a 0.5 mL UTEVA chromatographic resin bed and 2.5 mg of uranium.
Copyright © 2018 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Automation; Elemental Scientific; Nuclear; Spectroscopy; U(3)O(8); UTEVA

Year:  2018        PMID: 30086912     DOI: 10.1016/j.talanta.2018.06.063

Source DB:  PubMed          Journal:  Talanta        ISSN: 0039-9140            Impact factor:   6.057


  1 in total

1.  Determination of Pu-239 in Urine by Sector Field Inductively Coupled Plasma Mass Spectrometry (SF-ICP-MS) Using an Automated Offline Sample Preparation Technique.

Authors:  Ge Xiao; Robert L Jones
Journal:  J Radioanal Nucl Chem       Date:  2021-02       Impact factor: 1.371

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.