Literature DB >> 30086040

Biomechanical Assessment of Locking Plate Fixation of Comminuted Proximal Olecranon Fractures.

Amanda Fantry1, Andrew Sobel1, Nicholas Capito2, Jonathan D Hodax1, Tyler Pidgeon1, Sarath Koruprolu3, Ryan Rich3, Joseph L Fava4, Andrew Green1, Scott Paxton1.   

Abstract

OBJECTIVES: To determine if mean ultimate strength or failure mechanism differed between comminuted olecranon fractures created at the proximal 25% or 50% of the trochlear notch and fixed with precontoured posterior locking plates (PLPs).
METHODS: Comminuted osteotomies were created in 10 matched pairs of cadaveric upper extremities at either the proximal 25% or 50% of the trochlear notch after quantitative computed tomography scans were performed to evaluate bone mineral density. Variable-angle olecranon PLPs were fixed to the specimens. The triceps tendon of each specimen was loaded cyclically and then to failure. Comparison of mean force at failure (displacement >2 mm) was performed using the 2-tailed t test.
RESULTS: There were no significant differences in specimen bone mineral density within matched pairs. Nineteen specimens failed by olecranon bisection fracture in the sagittal plane. Specimens in the 25% osteotomy group failed at lower ultimate forces of 808 N (SD ± 474 N) versus 1058 N (SD ± 480 N) in the 50% osteotomy group (P = 0.044).
CONCLUSIONS: The ultimate strength of comminuted olecranon fracture fixation with a PLP decreases significantly if the fracture is proximal to the midpoint of the trochlear notch. Fractures proximal to the midpoint of the trochlear notch may benefit from supplemental fixation or suture augmentation to prevent failure, particularly at force ranges higher than those experienced during active elbow range of motion.

Entities:  

Mesh:

Year:  2018        PMID: 30086040     DOI: 10.1097/BOT.0000000000001279

Source DB:  PubMed          Journal:  J Orthop Trauma        ISSN: 0890-5339            Impact factor:   2.512


  3 in total

1.  Biomechanical comparison between double-plate fixation and posterior plate fixation for comminuted olecranon fracture using two triceps screws in synthetic bone model.

Authors:  Yohan Lee; Bong Wan Cho; Min Bom Kim; Young Ho Lee
Journal:  Medicine (Baltimore)       Date:  2022-01-07       Impact factor: 1.889

2.  Locked plating of geriatric olecranon fractures leads to low fixation failure and acceptable complication rates.

Authors:  Kelsey L Wise; Sarah Peck; Lauren Smith; Chad Myeroff
Journal:  JSES Int       Date:  2021-04-16

3.  Application of a novel shape-memory alloy concentrator in displaced olecranon fractures: a report of the technique and mid-term clinical results.

Authors:  Demeng Xia; Panyu Zhou; Lei Li; Yan Xia; Zichen Hao; Yuntong Zhang; Shuogui Xu
Journal:  J Orthop Surg Res       Date:  2020-10-02       Impact factor: 2.359

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.