Spring-actuated autoinjectors delivering viscous drug solutions resulting from large drug concentrations require large spring forces which can create high peak pressures and stresses within syringes. The high peak pressures and stresses can lead to device failure. Measurements with a suite of novel instrumentation and analysis using numerical simulation explain the peak pressures and peak stresses as originating from mechanical impacts between moving components, the large acceleration of the components, and surprisingly, the production of tension waves in the liquid resulting in cavitation. The presence and intensity of cavitation depend on relative timing between the pressurization and the acceleration of the syringe, which, in turn, depend on the size and location of an air gap inside the syringe. We show that production of localized but very high pressures can result from shock wave focusing in the conical section of the syringe.

Keywords:  Autoinjector; Cavitation; High-concentration drug solution; Pressure waves; Shock focusing; Stress waves; Viscous drug solution

Mesh:

Year:  2018        PMID: 30084014     DOI: 10.1007/s13346-018-0568-7

Source DB:  PubMed          Journal:  Drug Deliv Transl Res        ISSN: 2190-393X            Impact factor:   4.617


  2 in total

Review 1.  Challenges in the development of high protein concentration formulations.

Authors:  Steven J Shire; Zahra Shahrokh; Jun Liu
Journal:  J Pharm Sci       Date:  2004-06       Impact factor: 3.534

Review 2.  New autoinjector technology for the delivery of subcutaneous methotrexate in the treatment of rheumatoid arthritis.

Authors:  Michael Schiff; Jonathan Jaffe; Bruce Freundlich; Patrick Madsen
Journal:  Expert Rev Med Devices       Date:  2014-06-17       Impact factor: 3.166

  2 in total
  1 in total

1.  MFC: An open-source high-order multi-component, multi-phase, and multi-scale compressible flow solver.

Authors:  Spencer H Bryngelson; Kevin Schmidmayer; Vedran Coralic; Jomela C Meng; Kazuki Maeda; Tim Colonius
Journal:  Comput Phys Commun       Date:  2020-05-23       Impact factor: 4.717

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.