External electrical cardioversion in patients with cardiac implantable electronic devices: Is it safe and is immediate device interrogation necessary?

Nikki A H A Pluymaekers1, Elton A M P Dudink1, Lucas Boersma2, Ömer Erküner1, Marloes Gelissen1, Vincent van Dijk2, Maurits Wijffels2, Trang Dinh1, Kevin Vernooy1, Harry J Crijns1, Jippe Balt2, Justin G L M Luermans1.   

Abstract

BACKGROUND: Atrial tachyarrhythmias are common in patients with cardiac implantable electronic devices (CIEDs). Restoration of sinus rhythm by external electrical cardioversion (eECV) is frequently used to alleviate symptoms and to ensure optimal device function.
OBJECTIVES: To evaluate the safety of eECV in patients with contemporary CIEDs and to assess the need for immediate device interrogation after eECV.
METHODS: We conducted a retrospective observational study of 229 patients (27.9% female, age 69 ± 10 years) with a CIED (104 pacemakers, 69 implantable cardioverter defibrillators, and 56 biventricular devices) who underwent eECV between 2008 and 2016 in two centers. Data from device interrogation before eECV, immediately afterwards, and at first follow-up (FU) after eECV were collected. CIED-related complications and adverse events during and after eECV were recorded.
RESULTS: No significant differences between right atrial (RA) and right ventricular (RV) sensing or threshold values before eECV, immediately afterwards, or at FU were observed. A small yet significant decrease was observed in RA and RV impedance immediately after eECV (484 Ω vs 462 Ω, P < 0.001 and 536 Ω vs 514 Ω, P < 0.001, respectively). The RV impedance did not recover to the baseline value (538 Ω vs 527 Ω, P  =  0.02). The impedance changes were without clinical consequences. No changes in left ventricular lead threshold or impedance values were measured. No CIED-related complications or adverse events were documented following eECV.
CONCLUSION: eECV in patients with contemporary CIEDs is safe. There seems to be no need for immediate device interrogation after eECV.
© 2018 Wiley Periodicals, Inc.

Entities:  

Keywords:  CIED; external ECV; immediate device interrogation; safety

Mesh:

Year:  2018        PMID: 30080928     DOI: 10.1111/pace.13467

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

1.  External defibrillation resulting in loss of ventricular capture during atrial lead testing.

Authors:  Nicholas Palmeri; Andrew Locke; Patricia Tung
Journal:  HeartRhythm Case Rep       Date:  2021-11-27

2.  Cardioversion of atrial fibrillation and atrial flutter revisited: current evidence and practical guidance for a common procedure.

Authors:  Axel Brandes; Harry J G M Crijns; Michiel Rienstra; Paulus Kirchhof; Erik L Grove; Kenneth Bruun Pedersen; Isabelle C Van Gelder
Journal:  Europace       Date:  2020-08-01       Impact factor: 5.214

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.