Literature DB >> 30078572

Canal Transportation, Centering Ability, and Cyclic Fatigue Promoted by Twisted File Adaptive and Navigator EVO Instruments at Different Motions.

Pedro Marks Duarte1, Paula Barcellos da Silva1, Murilo Priori Alcalde2, Rodrigo Ricci Vivan2, Ricardo Abreu da Rosa3, Marco Antonio Húngaro Duarte2, Marcus Vinícius Reis Só1.   

Abstract

INTRODUCTION: This study compared the cyclic fatigue and the canal transportation promoted by Twisted File (TF) Adaptive and Navigator EVO systems when used with 2 different motions.
METHODS: Forty mesiobuccal roots of maxillary molars were scanned by using micro-computed tomography imaging before and after root canal preparation with the 2 instrument systems used with 2 motions (adaptive and continuous rotation). Samples were divided into 4 groups: TFA, TF Adaptive instruments under adaptive motion; TFC, TF Adaptive instruments under continuous motion; NA, Navigator instruments under adaptive motion; and NC, Navigator instruments under continuous motion. Root canals were prepared until 35.04 instruments. Apical transportation was analyzed by using micro-computed tomography at 3 levels: 3, 6, and 9 mm from the apex. The cyclic fatigue tests were performed by using a custom-made device. Ten instruments of each brand were activated by using a 6:1 reduction handpiece powered by a torque-controlled motor using the preset programs "custom mode" and "TF Adaptive" to activate 25.06 and 35.04 instruments. Kruskal-Wallis and Dunn tests were used to assess canal transportation, centering ability, and canal volume. The Student t test was used to evaluate cyclic fatigue (P = .05).
RESULTS: At 3 and 9 mm, the canal transportation and centering ability were similar in all groups (P > .05). At 6 mm, TFC presented higher canal transportation toward furcal region than NA and NC (P < .05). After canal preparation, TFA promoted great dentinal excision, presenting higher canal volume than NA and NC (P < .05). Higher cyclic fatigue resistance was observed under continuous than adaptive motion regardless of system or tip/diameter of the instrument (P < .05).
CONCLUSIONS: Both systems can be used under adaptive or continuous rotation. However, the life span of the instruments was higher when used under continuous rotation. Small canal transportation occurred when mesiobuccal root canals from maxillary molars were prepared until 35.04 instruments.
Copyright © 2018 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Adaptive movement; continuous rotation; cyclic fatigue; micro-computed tomographic imaging; root canal preparation; root canal transportation

Mesh:

Substances:

Year:  2018        PMID: 30078572     DOI: 10.1016/j.joen.2018.06.002

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  4 in total

1.  Mechanical Properties of ProTaper Gold, EdgeTaper Platinum, Flex Gold and Pro-T Rotary Systems.

Authors:  Murilo Alcalde; Marco Antonio Hungaro Duarte; Pablo Andrés Amoroso Silva; Pedro Henrique Souza Calefi; Emmanuel Silva; Jussaro Duque; Rodrigo Vivan
Journal:  Eur Endod J       Date:  2020-12

2.  Cyclic Fatigue Resistance of Nickel-Titanium Rotary Instruments after Simulated Clinical Use.

Authors:  Hana R Ubaed; Diyar Kh Bakr
Journal:  Appl Bionics Biomech       Date:  2022-10-11       Impact factor: 1.664

3.  Influence of glide path size and operating kinetics on time to reach working length and fracture resistance of Twisted File adaptive and Endostar E3 nickel-titanium file systems.

Authors:  Tamilkumaran Ramyadharshini; Inbaraj Anand Sherwood; V Shanmugham Vigneshwar; Prakasam Ernest Prince; Murugadoss Vaanjay
Journal:  Restor Dent Endod       Date:  2020-03-05

Review 4.  Present status and future directions: Canal shaping.

Authors:  Ana Arias; Ove A Peters
Journal:  Int Endod J       Date:  2022-02-16       Impact factor: 5.165

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.