Low Sensitivity of the Van Herick Method for Detecting Gonioscopic Angle Closure Independent of Observer Expertise.

Thomas V Johnson1, Pradeep Y Ramulu2, Harry A Quigley2, Eric L Singman2.   

Abstract

PURPOSE: To evaluate the diagnostic performance characteristics of the Van Herick assessment (VHA) for identifying angle closure compared to gold-standard gonioscopy, as conducted by trained observers of varying expertise.
DESIGN: Reliability analysis.
METHODS: Patients (n = 131) from a glaucoma referral clinic aged ≥50 years without prior ocular surgery or iridotomy underwent unilateral VHA by 1 of 11 trained ophthalmic technicians, followed by VHA and indirect gonioscopy by 1 of 15 ophthalmology residents and 1 of 4 glaucoma specialist attending physicians. Observers were masked to others' gradings. Cohen's kappa (κ) assessed test reproducibility. VHA sensitivity and specificity for identifying gonioscopic angle closure were calculated.
RESULTS: Mean patient age was 62.0 ± 8.7 years, 56% of patients were male, and 84% were African American. Angles were gonioscopically closed in 14.5% of eyes. Moderate agreement was observed comparing technician or resident VHA to attending VHA (κ = 0.48 and κ = 0.56, respectively). Resident and attending gonioscopy demonstrated excellent agreement (κ = 0.94). Sensitivities of technician, resident, and attending VHA for detecting angle closure were 57.9% (95% confidence interval: 34.0%-78.9%), 78.9% (53.9%-93.0%), and 68.4% (43.5%-86.4%), respectively. Specificities were 88.5% (80.3%-93.6%), 88.2% (80.3%-93.3%), and 87.5% (79.6%-92.8%), respectively.
CONCLUSIONS: VHA, even when performed by experienced ophthalmologists, misses a substantial proportion of angle closure while incorrectly identifying roughly 1 in 8 open-angle eyes as closed. These results suggest that clinical assessment of anterior chamber angle configuration is best accomplished with gonioscopy.
Copyright © 2018 Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2018        PMID: 30071210     DOI: 10.1016/j.ajo.2018.07.026

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  3 in total

1.  Potentially Missed Opportunities in Prevention of Acute Angle-Closure Crisis.

Authors:  Annie M Wu; Joshua D Stein; Manjool Shah
Journal:  JAMA Ophthalmol       Date:  2022-06-01       Impact factor: 8.253

2.  Non-contact tests for identifying people at risk of primary angle closure glaucoma.

Authors:  Anish Jindal; Irene Ctori; Gianni Virgili; Ersilia Lucenteforte; John G Lawrenson
Journal:  Cochrane Database Syst Rev       Date:  2020-05-28

3.  Diagnostic Performance of Optical Coherence Tomography and Nonspecialist Gonioscopy to Detect Angle Closure.

Authors:  Bruno Lb Esporcatte; Roberto M Vessani; Luiz As Melo; Norton S Yanagimori; Guilherme H Bufarah; Norma Allemann; Ivan M Tavares
Journal:  J Curr Glaucoma Pract       Date:  2022 Jan-Apr
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.