Multi-electrode Harmonized Kingdon Traps.

Evgeny Nikolaev1, Mikhail Sudakov2,3, Gleb Vladimirov2, Luis Fernando Velásquez-García4, Petr Borisovets2, Anastasia Fursova2.   

Abstract

Based on the analysis of the results of the study of various designs of multi-electrode harmonized Kingdon traps, we propose a new type of trap with two merged internal electrodes that has the ability to capture and accumulate ions formed inside. We also investigated the influence of inaccuracies in the manufacture of the electrodes on the field inside such trap. The four-electrode trap, which actually degenerates into a two-electrode device with traces of two other electrodes present at the ends of the internal electrodes (their splitting) has been found as the less sensitive to inaccuracies caused by manufacturing and cutting the ends of trap electrodes. We show that a mass spectrometer with a relatively high resolving power can be created on the basis of such a trap. The creation of the traps requires the manufacture of complex electrodes with demanded accuracy of their surfaces. This becomes possible with the advent of 3D printers. Graphical Abstract Multi electrode harmonize Kingdon trap with fused elctrode pares.

Keywords:  Electrostatic ion trap; FT-MS; Fourier transform; High resolution; Mass analyzer; Simulation

Year:  2018        PMID: 30069642     DOI: 10.1007/s13361-018-2032-9

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  1 in total

Review 1.  Evolution of Orbitrap Mass Spectrometry Instrumentation.

Authors:  Shannon Eliuk; Alexander Makarov
Journal:  Annu Rev Anal Chem (Palo Alto Calif)       Date:  2015       Impact factor: 10.745

  1 in total