| Literature DB >> 30061368 |
Troy D Long1, David F Kallmes2, Ricardo Hanel3, Tomoyoshi Shigematsu4, Alexander Michael Halaszyn5, Julia Wolter5, Alejandro Berenstein4.
Abstract
BACKGROUND: Navigable, large diameter aspiration catheters demonstrate markedly improved recanalization rates over smaller lumen devices in suction embolectomy. We evaluated the performance of a novel aspiration catheter system designed to maximize lumen size, and compare it to other commercially available aspiration catheters.Entities:
Keywords: brain; catheter; device; stroke; thrombectomy
Mesh:
Year: 2018 PMID: 30061368 PMCID: PMC6582808 DOI: 10.1136/neurintsurg-2017-013702
Source DB: PubMed Journal: J Neurointerv Surg ISSN: 1759-8478 Impact factor: 5.836
Figure 1Conceptual schematic of the R4Q friction fit into distal end of the guide catheter compared with the standard aspiration catheter system. The control wire (solid black) occupies less volume than the standard catheter within the guide catheter. Note: dimensions are not to scale.
Catheters
| Outer diameter | Catheter | Company | Distal ID (in) | Proximal ID (in) | Distal area (in2) | Length of tube (cm) | Effective ID (in) |
| 6F | R4Q 6F | MIVI Neuroscience, Inc. | 0.069 | 0.090* | 0.0037 | 115 | 0.0854 |
|
| |||||||
| Sophia Plus | MicroVention Terumo | 0.07 | 0.07 | 0.0038 | 131 | 0.0700 | |
|
| |||||||
| ACE 68 | Penumbra | 0.068 | 0.068 | 0.0036 | 132 | 0.0680 | |
|
| |||||||
| ACE 064 | Penumbra | 0.064 | 0.068 | 0.0032 | 132 | 0.0672 | |
|
| |||||||
| 5F | R4Q 5F | MIVI Neuroscience, Inc. | 0.057 | 0.090* | 0.0026 | 115 | 0.0828 |
|
| |||||||
| ARC | Medtronic | 0.061 | 0.069 | 0.0029 | 132 | 0.0682 | |
|
| |||||||
| Mi-Axus 6 | MIVI Neuroscience, Inc. | 0.057 | 0.064 | 0.0026 | 128 | 0.0633 | |
|
| |||||||
| InNeuroCo | Codman Neuro | 0.06 | 0.064 | 0.0028 | 125 | 0.0636 | |
|
| |||||||
| 5MAX ACE | Penumbra | 0.056 | 0.064 | 0.0028 | 132 | 0.0636 | |
|
| |||||||
| 5MAX | Penumbra | 0.054 | 0.064 | 0.0023 | 128 | 0.0630 | |
|
| |||||||
| 4F | R4Q 4F | MIVI Neuroscience, Inc. | 0.043 | 0.090* | 0.0015 | 120 | 0.0778 |
|
| |||||||
| 4MAX | Penumbra | 0.041 | 0.064 | 0.0013 | 139 | 0.0617 | |
|
| |||||||
| 3F | R4Q 3F | MIVI Neuroscience, Inc. | 0.036 | 0.090* | 0.00102 | 125 | 0.0768 |
|
| |||||||
| 3MAX | Penumbra | 0.035 | 0.043 | 0.00096 | 153 | 0.0422 | |
|
|
*Guide catheter ID is used as the proximal ID dimension of the R4Q system.
Figure 2Schematic of experimental setup for vacuum flow rate (A) and tip force measurements (B).
Flow rates
| Catheter | Flow rate (mL/s) |
| R4Q 6F | 5.31 |
| Sophia plus | 4.42 |
| ACE 068 | 4.27 |
| ACE 064 | 4.36 |
| R4Q 5F | 4.93 |
| ARC | 4.01 |
| Mi-Axus 6 | 3.82 |
| InNeuroCo 6F | 3.48 |
| 5MAX ACE | 3.60 |
| 5MAX | 3.00 |
| R4Q 4F | 3.89 |
| 4MAX | 2.44 |
| R4Q 3F | 3.50 |
| 3MAX | 1.03 |
Figure 3(A) Catheter flow rates by outer diameter size. (B) Average gram force measured for the ACE 68 and the R4Q.
Figure 4R4Q 6F flow rate (solid line) at multiple positions inside an 8F guide catheter compared with the constant flow rate of a similarly sized uniform catheter (dashed line).