| Literature DB >> 30060155 |
Maximilian Kreibich1,2, Matthias Siepe1,2, Friedhelm Beyersdorf1,2, Martin Czerny1,2.
Abstract
We report 2 cases of spontaneous, unexpected leakage of the stent graft portion of the Thoraflex (Vascutek, Inchinnan, UK) frozen elephant trunk prosthesis that occurred during second stage surgical thoraco-abdominal aortic replacement. When the descending aortas including the distal portion of the frozen elephant trunk grafts were clamped, spontaneous leakage occurred in the proximal, untouched stent graft portion. This brief communication highlights the previously unreported potential risk of spontaneous leakage through the stent graft portion of the frozen elephant trunk prosthesis during a second stage repair of the thoraco-abdominal aorta.Entities:
Mesh:
Year: 2019 PMID: 30060155 DOI: 10.1093/icvts/ivy238