Automated end-to-end blood testing at the point-of-care: Integration of robotic phlebotomy with downstream sample processing.

M L Balter1, J M Leipheimer1, A I Chen1, A Shrirao1, T J Maguire1, M L Yarmush1.   

Abstract

Diagnostic blood testing is the most commonly performed clinical procedure in the world, and influences the majority of medical decisions made in hospital and laboratory settings. However, manual blood draw success rates are dependent on clinician skill and patient physiology, and results are generated almost exclusively in centralized labs from large-volume samples using labor-intensive analytical techniques. This paper presents a medical device that enables end-to-end blood testing by performing blood draws and providing diagnostic results in a fully automated fashion at the point-of-care. The system couples an image-guided venipuncture robot, developed to address the challenges of routine venous access, with a centrifuge-based blood analyzer to obtain quantitative measurements of hematology. We first demonstrate a white blood cell assay on the analyzer, using a blood mimicking fluid spiked with fluorescent microbeads, where the area of the packed bead layer is correlated with the bead concentration. Next we perform experiments to evaluate the pumping efficiency of the sample handling module. Finally, studies are conducted on the integrated device - from blood draw to analysis - using blood vessel phantoms to assess the accuracy and repeatability of the resulting white blood cell assay.

Entities:  

Keywords:  Centrifugal Microfluidics; Medical Robotics; Microscale Sample Processing; Point-of-Care Blood Testing; Venous Access

Year:  2018        PMID: 30057935      PMCID: PMC6058193          DOI: 10.1142/S2339547818500048

Source DB:  PubMed          Journal:  Technology (Singap World Sci)


  24 in total

Review 1.  Commercialization of microfluidic point-of-care diagnostic devices.

Authors:  Curtis D Chin; Vincent Linder; Samuel K Sia
Journal:  Lab Chip       Date:  2012-02-17       Impact factor: 6.799

Review 2.  Laboratory automation: total and subtotal.

Authors:  Charles D Hawker
Journal:  Clin Lab Med       Date:  2007-12       Impact factor: 1.935

3.  Selecting automation for the clinical chemistry laboratory.

Authors:  Stacy E F Melanson; Neal I Lindeman; Petr Jarolim
Journal:  Arch Pathol Lab Med       Date:  2007-07       Impact factor: 5.534

4.  i-STAT - Combining Chemistry and Haematology in PoCT.

Authors:  Cameron L Martin
Journal:  Clin Biochem Rev       Date:  2010-08

5.  The difference between fingerstick and venous hemoglobin and hematocrit varies by sex and iron stores.

Authors:  Ritchard G Cable; Whitney R Steele; Russell S Melmed; Bryce Johnson; Alan E Mast; Patricia M Carey; Joseph E Kiss; Steven H Kleinman; David J Wright
Journal:  Transfusion       Date:  2011-10-20       Impact factor: 3.157

Review 6.  Difficult venous access in children: taking control.

Authors:  Laura L Kuensting; Scott DeBoer; Reneé Holleran; Barbara L Shultz; Rebecca A Steinmann; Jeanne Venella
Journal:  J Emerg Nurs       Date:  2009-03-21       Impact factor: 1.836

7.  Determination of turnaround time in the clinical laboratory: "accessioning-to-result" time does not always accurately reflect laboratory performance.

Authors:  Brie A Stotler; Alexander Kratz
Journal:  Am J Clin Pathol       Date:  2012-11       Impact factor: 2.493

8.  Assessment of using fingerstick blood sample with i-STAT point-of-care device for cardiac troponin I assay.

Authors:  Devin Loewenstein; Christine Stake; Mark Cichon
Journal:  Am J Emerg Med       Date:  2013-06-02       Impact factor: 2.469

9.  Evaluation of a new point of care automated complete blood count (CBC) analyzer in various clinical settings.

Authors:  L V Rao; Björn A Ekberg; Diane Connor; Felice Jakubiak; Guy M Vallaro; Michael Snyder
Journal:  Clin Chim Acta       Date:  2007-12-14       Impact factor: 3.786

10.  Point-of-care method for total white cell count: an evaluation of the HemoCue WBC device.

Authors:  A Osei-Bimpong; C Jury; R McLean; S M Lewis
Journal:  Int J Lab Hematol       Date:  2008-08-28       Impact factor: 2.877

View more
  5 in total

1.  Design and Evaluation of a Handheld Robotic Device for Peripheral Catheterization.

Authors:  Josh Leipheimer; Max Balter; Alvin Chen; Martin Yarmush
Journal:  J Med Device       Date:  2022-03-02       Impact factor: 0.743

2.  First-in-human evaluation of a hand-held automated venipuncture device for rapid venous blood draws.

Authors:  Josh M Leipheimer; Max L Balter; Alvin I Chen; Enrique J Pantin; Alexander E Davidovich; Kristen S Labazzo; Martin L Yarmush
Journal:  Technology (Singap World Sci)       Date:  2020-01-22

Review 3.  Anemia in salmon aquaculture: Scotland as a case study.

Authors:  A R Currie; D Cockerill; M Diez-Padrisa; H Haining; F L Henriquez; B Quinn
Journal:  Aquaculture       Date:  2022-01-15       Impact factor: 4.242

4.  InjectMeAI-Software Module of an Autonomous Injection Humanoid.

Authors:  Kwame Owusu Ampadu; Florian Rokohl; Safdar Mahmood; Marc Reichenbach; Michael Huebner
Journal:  Sensors (Basel)       Date:  2022-07-15       Impact factor: 3.847

Review 5.  Application of 5G network combined with AI robots in personalized nursing in China: A literature review.

Authors:  Caixia Guo; Hong Li
Journal:  Front Public Health       Date:  2022-08-24
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.