| Literature DB >> 30054130 |
Canan Armutcu1, Lokman Uzun1, Adil Denizli2.
Abstract
Automated on-line two-dimensional high-performance liquid chromatography method (2D-HPLC) is proposed to determine Ochratoxin A (OTA) in food samples as an alternative to OTA immunoaffinity column (IAC). An on-line 2D-HPLC system is designed for the analysis of OTA using an affinity-based monolithic column in the first dimension and reversed-phase C18 column in the second dimension. Initially, optimal OTA separation efficiency is determined through traditional HPLC system consisting of a P(HEMAPA) monolithic column coupled with HPLC system. Secondly, after providing optimum conditions, OTA determination was investigated through the 2D-HPLC system. According to results, 2D-HPLC system showed good linearity in the range 0.5 to 20 ng/mL with limit of detection (LOD) and limit of quantification (LOQ) values of 21.2 pg/mL and 64.3 pg/mL, respectively. The P(HEMAPA)-4 monolithic column displayed good recovery of OTA ranging from 104.34% to 107.33%. Relative standard deviations (RSD) varied in the range 0.21% to 1.31% thus indicating the efficiency of P(HEMAPA)-4 monolithic column developed for OTA.Entities:
Keywords: Food samples; Monolithic column; Ochratoxin A; Two-dimensional high-performance liquid chromatography
Mesh:
Substances:
Year: 2018 PMID: 30054130 DOI: 10.1016/j.chroma.2018.07.057
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759