Rapid qualitative and quantitative analysis of methamphetamine, ketamine, heroin, and cocaine by near-infrared spectroscopy.

Cui-Mei Liu1, Yu Han2, Shun-Geng Min3, Wei Jia2, Xin Meng2, Pei-Pei Liu2.   

Abstract

Rapid and nondestructive near infrared spectroscopy (NIR) methods have been developed for simultaneous qualitative and quantitative analysis of methamphetamine, ketamine, heroin, and cocaine in seized samples. This is the first systematic report regarding a qualitative and quantitative procedure of applying NIR for drug analysis. A total of 282 calibration samples and 836 prediction samples were used for the building and validating of qualitative and quantitative models. Two qualitative analysis modeling methods for soft independent modeling by class analogy (SIMCA) and supporting vector machine (SVM) were compared. From its excellent performance in rejecting false positive results, SIMCA was chosen. The drug concentrations in the calibration and validation sample sets were analyzed using high-performance liquid chromatography. Based on the use of first-order derivative spectral data after standard normal variate (SNV) transformation correction, in the wavelength range from 10,000 to 4000cm-1, four partial least squares quantitative-analysis models were built. The coefficients of determination for all calibration models were >99.3, and the RMSEC, RMSECV, and RMSEP were all less than 1.6, 2.9, and 3.6%, respectively. The results obtained here indicated that NIR with chemometric methods was accurate for qualitative and quantitative analysis of drug samples. This methodology provided a potentially useful alternative to time-consuming gas chromatography-mass spectroscopy and high-performance liquid chromatography methods.
Copyright © 2018 Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Near infrared spectroscopy; Partial least squares; Qualitative and quantitative analysis; SIMCA; drugs

Mesh:

Substances:

Year:  2018        PMID: 30053735     DOI: 10.1016/j.forsciint.2018.07.008

Source DB:  PubMed          Journal:  Forensic Sci Int        ISSN: 0379-0738            Impact factor:   2.395


  5 in total

Review 1.  Interpol review of controlled substances 2016-2019.

Authors:  Nicole S Jones; Jeffrey H Comparin
Journal:  Forensic Sci Int Synerg       Date:  2020-05-24

2.  Increased Amplitude of Low-Frequency Fluctuation in Right Angular Gyrus and Left Superior Occipital Gyrus Negatively Correlated With Heroin Use.

Authors:  Jing Luo; Ru Yang; Wenhan Yang; Chunmei Duan; Yuan Deng; Jun Zhang; Jiyuan Chen; Jun Liu
Journal:  Front Psychiatry       Date:  2020-07-03       Impact factor: 4.157

3.  Rapid and robust on-scene detection of cocaine in street samples using a handheld near-infrared spectrometer and machine learning algorithms.

Authors:  Ruben F Kranenburg; Joshka Verduin; Yannick Weesepoel; Martin Alewijn; Marcel Heerschop; Ger Koomen; Peter Keizers; Frank Bakker; Fionn Wallace; Annette van Esch; Annemieke Hulsbergen; Arian C van Asten
Journal:  Drug Test Anal       Date:  2020-07-27       Impact factor: 3.345

Review 4.  Research Progress of Applying Infrared Spectroscopy Technology for Detection of Toxic and Harmful Substances in Food.

Authors:  Wenliang Qi; Yanlong Tian; Daoli Lu; Bin Chen
Journal:  Foods       Date:  2022-03-23

5.  A calibration friendly approach to identify drugs of abuse mixtures with a portable near-infrared analyzer.

Authors:  Ruben F Kranenburg; Henk-Jan Ramaker; Sharon Sap; Arian C van Asten
Journal:  Drug Test Anal       Date:  2022-02-09       Impact factor: 3.234