Literature DB >> 30037958

Mural destabilization after aneurysm treatment with a flow-diverting device: a report of two cases.

Timothy Hampton1, Donal Walsh1, Christos Tolias1, David Fiorella2.   

Abstract

BACKGROUND: Flow-diverting stents have demonstrated great promise for the treatment of cerebral aneurysms; however, clinical experience with the devices remains very preliminary. We present two cases of spontaneous delayed complications-one fatal aneurysm rupture and one symptomatic increase in aneurysm volume-following the treatment of intradural aneurysms with the Pipeline Embolization Device (PED). PRESENTATION/INTERVENTION: Two patients with unruptured, intradural aneurysms of the carotid artery underwent uneventful treatment with the PED (eV3, Irvine, California, USA). One patient, with a giant aneurysm of the carotid terminus, experienced worsening headache 5 days after the procedure and ultimately collapsed and became unresponsive. CT of the head demonstrated acute subarachnoid and intraventricular hemorrhage. The patient died the following day. A second patient with a large left posterior communicating artery aneurysm presented with progressive memory loss 3 months after PED reconstruction of the carotid artery. Although serial CT angiograms showed progressive thrombosis of the aneurysm to near-complete occlusion, MR of the brain demonstrated marked interval growth of the collective aneurysm-intra-aneurysmal thrombus mass with extensive edema throughout the adjacent left temporal lobe.
CONCLUSIONS: Flow-diverting devices have demonstrated tremendous promise for the treatment of complex, unruptured cerebral aneurysms. However, experience with this novel approach to aneurysm treatment is preliminary and the consequences of its application within the cerebrovasculature remain incompletely defined. Mural destabilization resulting in delayed, spontaneous, aneurysm growth and/or rupture may occur in the days to weeks following the application of flow-diverting devices to treat previously unruptured intracranial aneurysms. A better understanding of the incidence and etiology of these complications is essential for this technology to be optimally applied.
© 2011, Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://group.bmj.com/group/rights-licensing/permissions.

Entities:  

Keywords:  Aneurysm; complication; device; hemorrhage; pipeline embolization device; spontaneous rupture; stent

Mesh:

Substances:

Year:  2018        PMID: 30037958     DOI: 10.1136/jnis.2010.002873.rep

Source DB:  PubMed          Journal:  J Neurointerv Surg        ISSN: 1759-8478            Impact factor:   5.836


  6 in total

1.  Current Strategies in the Treatment of Intracranial Large and Giant Aneurysms.

Authors:  Matthias Gmeiner; Andreas Gruber
Journal:  Acta Neurochir Suppl       Date:  2021

2.  Outcome of flow diverter placement for intracranial aneurysm with dual antiplatelet therapy and oral anticoagulant therapy.

Authors:  Takashi Fujii; Hidenori Oishi; Kohsuke Teranishi; Kenji Yatomi; Kazumoto Suzuki; Hajime Arai
Journal:  Interv Neuroradiol       Date:  2020-07-29       Impact factor: 1.610

3.  Delayed rupture of intracranial aneurysms after placement of intra-luminal flow diverter.

Authors:  Kun Hou; Guichen Li; Xianli Lv; Baofeng Xu; Kan Xu; Jinlu Yu
Journal:  Neuroradiol J       Date:  2020-08-27

4.  Hemodynamic Analysis of Postoperative Rupture of Unruptured Intracranial Aneurysms after Placement of Flow-Diverting Stents: A Matched Case-Control Study.

Authors:  W Li; Z Tian; W Zhu; Y S Zhang; K Wang; Y Zhang; Y Wang; X Yang; J Liu
Journal:  AJNR Am J Neuroradiol       Date:  2019-10-17       Impact factor: 3.825

5.  Delayed spontaneous rupture of cavernous segment of the internal carotid artery following dual ophthalmic segment aneurysms treatment with pipeline embolization device: A case report.

Authors:  Wang Ting; Seidu A Richard; Zhang Changwei; Wang Chaohua; Xie Xiaodong
Journal:  Medicine (Baltimore)       Date:  2019-12       Impact factor: 1.817

6.  Endovascular Treatment of Large or Giant Basilar Artery Aneurysms Using the Pipeline Embolization Device: Complications and Outcomes.

Authors:  Huijian Ge; Xiheng Chen; Kai Liu; Yang Zhao; Longhui Zhang; Peng Liu; Yuhua Jiang; Hongwei He; Ming Lv; Youxiang Li
Journal:  Front Neurol       Date:  2022-03-02       Impact factor: 4.003

  6 in total