Samantha M Govender1, Maurice Mars.
Abstract
BACKGROUND: Asynchronous automated telehealth-based hearing screening and diagnostic testing can be used within the rural school context to identify and confirm hearing loss. Objective: The aims of the study were to evaluate the efficacy of an asynchronous telehealth-based service delivery model using automated technology for screening and diagnostic testing as well as to describe the prevalence, type and degree of hearing loss. Method: A comparative within-subject design was used. Frequency distributions, sensitivity, specificity scores as well as the positive and negative predictive values were calculated. Testing was conducted in a non-sound-treated classroom within a school environment on 73 participants (146 ears). The sensitivity and specificity rates were 65.2% and 100%, respectively. Diagnostic accuracy was 91.7% and the negative predictive values (NPV) and positive predictive values (PPV) were 93.8% and 100%, respectively.Entities:
Keywords: automated audiometry; diagnostic audiometry; referral; screening; sensitivity; specificity; telehealth
Mesh:
Year: 2018 PMID: 30035608 PMCID: PMC6111388 DOI: 10.4102/sajcd.v65i1.582
Source DB: PubMed Journal: S Afr J Commun Disord ISSN: 0379-8046
FIGURE 1Clinical protocol.
Outcome of screening and diagnostic tests.
| Variable | Diagnostic test positive | Diagnostic test negative | % | 95% confidence intervals |
|---|---|---|---|---|
| Screening test (positive) | 15 | 0 | - | - |
| Screening test (negative) | 8 | 123 | - | - |
| Sensitivity | - | - | 65.2 | 42.77% – 83.61% |
| Specificity | - | - | 100 | 97.04% – 100% |
| Positive predictive value | - | - | 100 | 78.19% – 100% |
| Negative predictive value | - | - | 93.9 | 88.31% – 96.42% |
| Diagnostic accuracy | - | - | 94.5 | 89.7% – 97.33% |
| Disease prevalence | - | - | 15.7 | 10.26% – 22.69% |
Thresholds across frequency ranges for diagnostic (D) and screening (S) tests for eight false-negative screening tests.
| Number | 125 Hz | 250 Hz | 500 Hz | 1000 Hz | 2000 Hz | 4000 Hz | 8000 Hz | ||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| D | D | D | S | D | S | D | S | D | S | D | |
| 1 | 25 | 25 | 25 | 25 | 25 | 25 | 25 | ||||
| 2 | 15 | 25 | 20 | 25 | 20 | ||||||
| 3 | |||||||||||
| 4 | 20 | 25 | 10 | 25 | 5 | ||||||
| 5 | 15 | 25 | 10 | 25 | 10 | ||||||
| 6 | 25 | 25 | 25 | 25 | 15 | ||||||
| 7 | 25 | 25 | 20 | 25 | 15 | 25 | 15 | ||||
| 8 | 25 | 25 | 25 | 25 | 10 | 25 | 10 | ||||
Note: Discrepancies between the tests are in bold.