Evaluating the analytical performance of four new coagulation assays for the measurement of fibrinogen, D-dimer and thrombin time.

Steve Kitchen1, Ulrich Geisen2, Janos Kappelmayer3, Peter Quehenberger4, Anna Lowe1, Robert Jones1, Gabrielle Miles5, Jana G Boehm6, Gergely Rozsnyai7.   

Abstract

INTRODUCTION: New laboratory methods to measure haemostatic function require careful assessment before routine use. We evaluated the analytical performance of four new coagulation assays for the measurement of fibrinogen by Clauss assay, prothrombin time-derived fibrinogen, thrombin time and D-dimer levels.
METHODS: The four assays were evaluated on the cobas t 711 and cobas t 511 analysers at four centres in Europe. Analytical performance and method comparisons with other commercially available assays were performed according to Clinical and Laboratory Standards Institute guidelines (EP09-A3, EP05-A3) using residual anonymized human sodium citrate (3.2% [0.109M]) plasma samples. Lot-to-lot variability and the equivalency of each assay on the cobas t 711 and cobas t 511 analysers were also assessed.
RESULTS: Overall, coefficients of variance were ≤4.1% and ≤8.6% for within-run precision and total reproducibility, respectively. Method comparison experiments showed good or acceptable agreement for each assay compared with their respective comparator method, and equivalency was demonstrated for the two cobas t platforms (Pearson's correlation coefficient ≥0.991). A high level of consistency was observed between lots for all four assays (Pearson's correlation coefficient ≥0.994).
CONCLUSION: This multicentre study demonstrates excellent analytical performance for four new coagulation assays on the cobas t 711 and cobas t 511 analysers.
© 2018 The Authors. International Journal of Laboratory Hematology Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  D-dimer; cobas t 511; cobas t 711; fibrinogen; thrombin time

Mesh:

Substances:

Year:  2018        PMID: 30033683     DOI: 10.1111/ijlh.12896

Source DB:  PubMed          Journal:  Int J Lab Hematol        ISSN: 1751-5521            Impact factor:   2.877


  5 in total

1.  Activated clotting time in inpatient diagnostic and interventional settings.

Authors:  Heidi Dalton; Michael Martin; Pamela Garcia-Filion; David Shavelle; Pei-Hsiu Huang; Justin Clark; Sean Beinart; Andrew Israel; Nichole Korpi-Steiner
Journal:  J Thromb Thrombolysis       Date:  2022-06-22       Impact factor: 2.300

2.  Assessment and establishment of a reference interval for Roche Cobas t 711 coagulation analyzer for a hospital in China.

Authors:  Chi Zhu; Lili Sun; Haowei Li; Zhenghua Dong; Siyuan Yu; Xiaoming Zhao; Ji Yang; Wenjuan Wu
Journal:  Pract Lab Med       Date:  2022-02-17

3.  Comparison of Fibrinogen Concentrations Determined by the Clauss Method with Prothrombin-Derived Measurements on an Automated Coagulometer.

Authors:  Berrak Guven; Murat Can; Abdulkadir Tekin
Journal:  J Appl Lab Med       Date:  2022-08-22

4.  Performance of 6 routine coagulation assays on the new Roche Cobas t711 analyzer.

Authors:  Marlies Oostendorp; Roefke L Noorman; J Dinant Nijenhuis; Jacques B de Kok
Journal:  Pract Lab Med       Date:  2019-11-09

5.  System performance evaluation of the cobas t 711 and cobas t 511 coagulation analyzers in routine laboratory settings.

Authors:  Steve Kitchen; Moniek de Maat; Michael Nagler; Robert Jones; Anna Lowe; John Burden; Kai Groneberg; Gergely Rozsnyai
Journal:  Blood Coagul Fibrinolysis       Date:  2020-10       Impact factor: 1.061

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.