| Literature DB >> 30033672 |
Anita Berndt1, Monique van Prooijen2, Mathieu Guillot3.
Abstract
The Canadian Organization of Medical Physicists (COMP), in close partnership with the Canadian Partnership for Quality Radiotherapy (CPQR), has developed a series of Technical Quality Control (TQC) guidelines for radiation treatment equipment. These guidelines outline the performance objectives that equipment should meet in order to ensure an acceptable level of radiation treatment quality. The TQC guidelines have been rigorously reviewed and field tested in a variety of Canadian radiation treatment facilities. The development process enables rapid review and update to keep the guidelines current with changes in technology (the most updated version of this guideline can be found on the CPQR website). This particular TQC details recommended quality control testing for Gamma Knife radiosurgery.Entities:
Keywords: Gamma Knife; quality assurance; radiosurgery
Mesh:
Year: 2018 PMID: 30033672 PMCID: PMC6123156 DOI: 10.1002/acm2.12406
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Daily/weekly quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Daily | |||
| D1 | GK unit interlocks (frame adapter, side panels) | Functional | |
| D2 | Timer accuracy, linearity | 1%, 0.5% | 2%, 1% |
| D3 | Treatment console alarm test | Functional | |
| D4 | Emergency procedure placards | Present | |
| Weekly | |||
| W1 | Focus precision test | Functional | |
Monthly/quarterly quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Monthly | |||
| M1 | Clearance test tool check | Functional | |
| M2 | UPS battery check | Functional | |
| M3 | Patient positioning system retraction | Functional | |
| M4 | Patient positioning system accuracy | n/a | 0.5 mm |
| Quarterly | |||
| Q1 | Sector alignment | n/a | 0.5/1.0 mm |
0.5 mm for 4 and 8 mm collimators, 1.0 mm for 16 mm collimator.
Annual quality control tests
| Designator | Test | Performance | |
|---|---|---|---|
| Tolerance | Action | ||
| Annual | |||
| A1 | Coincidence of radiation and mechanical isocenter | 0.1 mm relative to baseline | 0.5 mm absolute |
| A2 | Timer linearity | 0.5% | 1% |
| A3 | Timer transit error | Baseline | |
| A4 | Profile accuracy | n/a | 1 mm |
| A5 | Backup timer on GK sector computer | Functional | |
| A6 | Absolute calibration | 1% | 2% |
| A7 | External service dose verification | n/a | 5% |
| A8 | End‐to‐end test | 1–5%/0.5 mm | 1–5%/1.0 mm |
| A9 | Radiation leak test | Baseline | |
| A10 | Radiation survey | Background | |
| A11 | Independent quality control review | Complete | |
After each major maintenance and then every other year; tolerance as per testing institution (e.g., Imaging and Radiation Oncology Core [IROC]).
| D1 | The GK inhibits beam on if the patient is not locked in place, at the correct gamma angle with the side protection panels engaged |
| D2 | The GK timer agrees with an independent measurement (e.g., stopwatch). Linearity can be tested by cycling through shots of different durations over multiple days |
| D3 | The GK built‐in alarm test causes the console alarm to sound |
| D4 | The emergency procedure placards are posted |
| W1 | The GK built‐in focus precision test indicates “PASS” |
| M1 | The GK clearance test tool check passes. Also check after possible damage to the tool. At the discretion of the physicist, test frequency may be reduced to semiannually |
| M2 | The Elekta uninterruptible power supply (UPS) check passes |
| M3 | Disengaging the x/z couch clutch allows the couch to be manually moved in the x/z direction |
| M4 | The position of the patient positioning system must be verified against physical reference positions over an appropriate clinical range in the directions of the three axes (x,y,z) |
| Q1 | The sectors move to correct alignment with the 4, 8, or 16 mm collimators |
| A1 | The positions of the radiation and mechanical isocenters agree with each other |
| A2 | The GK timer is linear. Test over a larger range than daily testing |
| A3 | The transit error is consistent with that measured during commissioning |
| A4 | The measured profiles agree with those in the treatment planning system. Stated tolerance applies to the 50% isodose line for each collimator size |
| A5 | The backup timer on the GK sector computer agrees with the console computer |
| A6 | The absolute dose rate in the treatment planning system matches the measured dose rate. Measurements must be made with a calibrated chamber using an accepted protocol (e.g., TG‐21 |
| A7 | The absolute dose is independently verified by an external service (e.g., IROC Houston OSLD/TLD [optically‐stimulated/thermoluminescent dosimeter] Monitoring Program) |
| A8 | An end‐to‐end phantom test is performed including frame placement, imaging, treatment planning, treatment, and verification that the intended treatment was delivered with the stated dose and positioning accuracy. The dosimetric accuracy depends on the dosimeter being used. For example, 1% accuracy would apply when using an ion chamber, whereas 5% would be appropriate for film |
| A9–10 | The configuration of these tests will depend on the design of the facility and equipment. As a minimum, Canadian Nuclear Safety Commission (CNSC) license conditions and applicable regulations must be followed |
| A11 | To ensure redundancy and adequate monitoring, a second qualified medical physicist must independently verify the implementation, analysis, and interpretation of the quality control tests at least annually |