Force and Contact Location Measurement Errors of the VERASENSE.

Stephanie J Nicolet-Petersen1, Stephen M Howell2, Maury Hull3.   

Abstract

BACKGROUND: The OrthoSensor VERASENSE knee system is a commercially available instrumented tibial insert that provides real-time intraoperative measurements of tibial contact force and contact location to guide surgeons toward improving outcomes in total knee arthroplasty (TKA). However, the device has been used contrary to the manufacturer's recommendations in several studies and lacks published accuracy data. Therefore, the primary objectives of this study were to evaluate the device's error in tibial contact force when used according to and contrary to the manufacturer's recommendations, and also to evaluate the device's error in anterior-posterior (A-P) and medial-lateral (M-L) contact locations.
METHODS: The error in tibial contact force in single compartment distributed loading was evaluated by applying known forces in ranges within and exceeding that recommended by the manufacturer, with rezeroing as recommended by the manufacturer, and without rezeroing. The error in tibial contact location in single compartment concentrated loading was evaluated by applying known forces at known locations on the articular surface.
RESULTS: Exceeding the maximum allowable load and not rezeroing did not adversely affect the bias (i.e. average error) (p > 0.05). The maximum absolute bias without rezeroing was 2.9 lbf. Rezeroing more than doubled the bias. The maximum root mean squared error in tibial contact location was 1.5 mm in the A-P direction.
CONCLUSION: The device measures tibial contact force with comparable error well above the maximum allowable load and without rezeroing, contrary to the manufacturer's instructions.

Year:  2018        PMID: 30029259     DOI: 10.1115/1.4040601

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  3 in total

1.  Variability between the trial and final implant measurements during the sensor-guided total knee arthroplasty.

Authors:  Jong-Keun Kim; Do Weon Lee; Du Hyun Ro; Hyuk-Soo Han; Myung Chul Lee
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2022-01-04       Impact factor: 4.114

2.  What is the Effect of Posterior Osteophytes on Flexion and Extension Gaps in Total Knee Arthroplasty? A Cadaveric Study.

Authors:  David C Holst; Gary W Doan; Marc R Angerame; Martin W Roche; Chadd W Clary; Douglas A Dennis
Journal:  Arthroplast Today       Date:  2021-09-06

3.  Intraoperative Load Sensing in Total Knee Arthroplasty Leads to a Functional but Not Clinical Difference: A Comparative, Gait Analysis Evaluation.

Authors:  Michele Giuntoli; Michelangelo Scaglione; Enrico Bonicoli; Nicola Piolanti; Gianmarco Puccioni; Karlos Zepeda; Emanuele Giannini; Stefano Marchetti; Pier Francesco Indelli
Journal:  J Funct Morphol Kinesiol       Date:  2022-02-18
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.