Literature DB >> 30029242

Laboratory Validation of a Wearable Sensor for the Measurement of Head Acceleration in Men's and Women's Lacrosse.

Jessica M Buice1, Amanda O Esquivel2, Christopher J Andrecovich3.   

Abstract

Mild traumatic brain injuries, or concussions, can result from head acceleration during sports. Wearable sensors like the GForceTrackerTM (GFT) can monitor an athlete's head acceleration during play. The purpose of this study was to evaluate the accuracy of the GFT for use in boys' and girls' lacrosse. The GFT was mounted to either a strap connected to lacrosse goggles (helmetless) or a helmet. The assembly was fit to a Hybrid III (HIII) headform instrumented with sensors and impacted multiple times at different velocities and locations. Measurements of peak linear acceleration and angular velocity were obtained from both systems and compared. It was found that a large percent error between the GFT and headform system existed for linear acceleration (29% for helmetless and 123% for helmet) and angular velocity (48% for helmetless and 17% for helmet). Linear acceleration data transformed to the center of gravity (CG) of the head still produced errors (47% for helmetless and 76% for helmet). This error was substantially reduced when correction equations were applied based on impact location (3-22% for helmetless and 3-12% for helmet impacts at the GFT location and transformed to the CG of the head). Our study has shown that the GFT does not accurately calculate linear acceleration or angular velocity at the CG of the head; however, reasonable error can be achieved by correcting data based on impact location.

Entities:  

Mesh:

Year:  2018        PMID: 30029242     DOI: 10.1115/1.4040311

Source DB:  PubMed          Journal:  J Biomech Eng        ISSN: 0148-0731            Impact factor:   2.097


  2 in total

1.  Head Impact Research Using Inertial Sensors in Sport: A Systematic Review of Methods, Demographics, and Factors Contributing to Exposure.

Authors:  Enora Le Flao; Gunter P Siegmund; Robert Borotkanics
Journal:  Sports Med       Date:  2021-10-22       Impact factor: 11.136

Review 2.  The Use of Wearable Sensors in Human Movement Analysis in Non-Swimming Aquatic Activities: A Systematic Review.

Authors:  Daniel A Marinho; Henrique P Neiva; Jorge E Morais
Journal:  Int J Environ Res Public Health       Date:  2019-12-12       Impact factor: 3.390

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.