Linearity check of an ionisation chamber through 99 mTc half-life measurements.

S Pommé1, J Paepen2, R Van Ammel2.   

Abstract

The half-life of 99 mTc was measured at the JRC using the ionisation chamber 'IC1' (type Centronic IG12). The result, T1/2(99 mTc) = 6.00660 (18) h, is in good agreement with literature data. One experiment was performed in IC1's default set-up with the ionisation current being integrated over an air capacitor and read out as a voltage increase over time. This ensured excellent linearity and precision throughout the dynamic range, but the maximum current was limited to 2 nA. In a second test, the current was directly read out with a Keithley 6517 A electrometer. Applying correction factors for the automatic range switching of the electrometer, an acceptable linearity was demonstrated over a range of 12 half-life periods starting at 20 nA. Range switching and autocorrelation of the current readout increase the systematic and random error propagation factors. Piecewise fitting of the decay curve over periods of 6 h yields the same 99 mTc half-life value within 0.04% (0.0025 h) standard deviation over an activity range spanning at least 10 half-life periods (3 orders of magnitude).
Copyright © 2018 The Authors. Published by Elsevier Ltd.. All rights reserved.

Entities:  

Keywords:  (99 m)Tc; Decay constant; Half-life; Ionisation chamber; Linearity; Uncertainty

Year:  2018        PMID: 30029094     DOI: 10.1016/j.apradiso.2018.07.004

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  1 in total

1.  Air humidity and annual oscillations in 90Sr/90Y and 60Co decay rate measurements.

Authors:  S Pommé; K Pelczar; I Kajan
Journal:  Sci Rep       Date:  2022-06-09       Impact factor: 4.996

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.