| Literature DB >> 30028688 |
Michael Rapin, Fabian Braun, Andy Adler, Josias Wacker, Inez Frerichs, Barbara Vogt, Olivier Chetelat.
Abstract
This paper presents a wearable sensor architecture for frequency-multiplexed electrical impedance tomography (EIT) and synchronous multilead electrocardiogram (ECG) data acquisition. The system is based on a novel electronic sensing architecture, called cooperative sensors, that significantly reduces the cabling complexity and enables flexible EIT stimulation and measurement patterns. The cooperative-sensor architecture was initially designed for ECG and has been extended for multichannel bioimpedance measurement. This approach allows for an adjustable EIT stimulation pattern via frequency-division multiplexing. This paper also shows a calibration procedure as well as EIT system noise performance assessment. Preliminary measurements on a healthy volunteer showed the ability of the wearable system to measure EIT data synchronously with multilead ECG. Ventilation-related and heartbeat-related EIT images were reconstructed, demonstrating the feasibility of the proposed architecture for noninvasive cardiovascular monitoring.Mesh:
Year: 2018 PMID: 30028688 DOI: 10.1109/TBME.2018.2857199
Source DB: PubMed Journal: IEEE Trans Biomed Eng ISSN: 0018-9294 Impact factor: 4.538