Literature DB >> 30026256

Hydrophilic polymer embolic complication during diagnostic cerebral angiography presenting with delayed intracranial hemorrhage: case report and literature review.

Sameer A Ansari1,2,3, Richard R Anderson4, Michael J Caron3,5, Ali Shaibani1,3, Michael C Hurley1,3, Babak S Jahromi1,3, Matthew B Potts1,3.   

Abstract

We report two serial neuroendovascular cases of hydrophilic polymer embolic complications, and highlight a unique case of a routine diagnostic cerebral angiogram that was complicated by delayed intracranial hemorrhage requiring surgical decompression. Histopathology specimens revealed organized intravascular thrombi with foci of non-polarizable, basophilic foreign material. Shavings from the hydrophilic coatings of a standard diagnostic catheter and guidewire share histologic characteristics with this intravascular foreign material, confirming the diagnosis of hydrophilic polymer emboli. While this phenomenon has been described for complex neurointerventional procedures, it is rare with routine diagnostic cerebral angiography. Along with a detailed literature review, these cases provides further evidence that even basic hydrophilic coated catheters and/or wires may contribute to the etiology of iatrogenic emboli in the neurovasculature with the potential for acute and subacute complications, requiring further investigation. © Author(s) (or their employer(s)) 2019. No commercial re-use. See rights and permissions. Published by BMJ.

Entities:  

Keywords:  catheter; complication; device; embolic; material

Mesh:

Substances:

Year:  2018        PMID: 30026256     DOI: 10.1136/neurintsurg-2018-014189

Source DB:  PubMed          Journal:  J Neurointerv Surg        ISSN: 1759-8478            Impact factor:   5.836


  4 in total

1.  Delayed Leukoencephalopathy: A Rare Complication after Coiling of Cerebral Aneurysms.

Authors:  A Ikemura; T Ishibashi; K Otani; I Yuki; T Kodama; I Kan; N Kato; Y Murayama
Journal:  AJNR Am J Neuroradiol       Date:  2020-01-30       Impact factor: 3.825

2.  Suspected Metallic Embolization Distal to Coiled Intracranial Aneurysms Detectable by Susceptibility-Weighted MR Imaging.

Authors:  D H Yoo; Y D Cho; H S Lee; S H Kim; D Jang; S H Lee; W-S Cho; H-S Kang; J E Kim; M H Han
Journal:  AJNR Am J Neuroradiol       Date:  2020-04       Impact factor: 3.825

3.  Polytetrafluoroethylene coating fragments during neuroendovascular therapy: An analysis of two damaged microguidewires.

Authors:  Rasmus Holmboe Dahl; René Wugt Larsen; Esben Thormann; Goetz Benndorf
Journal:  Interv Neuroradiol       Date:  2021-05-26       Impact factor: 1.610

4.  Intrathrombus polymer coating deposition: a pilot study of 91 patients undergoing endovascular therapy for acute large vessel stroke. Part I: Histologic frequency.

Authors:  Rashi I Mehta; Ansaar T Rai; Jeffrey A Vos; Orestes E Solis; Rupal I Mehta
Journal:  J Neurointerv Surg       Date:  2019-05-18       Impact factor: 5.836

  4 in total