Literature DB >> 30019453

Entrapment of "PV-Tracker" guidewire in the pulmonary veins during atrial fibrillation ablation with PVAC Gold.

Adolfo Fontenla1, María López-Gil1, Isabel Montilla1, Fernando Arribas1.   

Abstract

We report the entrapment of Pulmonary Vein Ablation Catheter (Medtronic, Minneapolis, MN, USA) and its guidewire within the right pulmonary veins in two patients. The catheters could be retrieved without complications but they were disabled in both cases. This nondescribed incident should be recognized by PVAC users since it may be a source of potential severe complications during pulmonary vein isolation procedures.
© 2018 Wiley Periodicals, Inc.

Entities:  

Keywords:  Pulmonary Vein Ablation Catheter (PVAC); atrial fibrillation ablation; catheter entrapment; phased radiofrequency; pulmonary veins isolation

Mesh:

Year:  2018        PMID: 30019453     DOI: 10.1111/pace.13451

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  1 in total

1.  Entrapment of a spline from a multielectrode mapping catheter in the side hole of a nonsteerable guiding sheath, with dislodgement and subsequent suction retrieval of a proximal spline electrode.