| Literature DB >> 30019453 |
Adolfo Fontenla1, María López-Gil1, Isabel Montilla1, Fernando Arribas1.
Abstract
We report the entrapment of Pulmonary Vein Ablation Catheter (Medtronic, Minneapolis, MN, USA) and its guidewire within the right pulmonary veins in two patients. The catheters could be retrieved without complications but they were disabled in both cases. This nondescribed incident should be recognized by PVAC users since it may be a source of potential severe complications during pulmonary vein isolation procedures.Entities:
Keywords: Pulmonary Vein Ablation Catheter (PVAC); atrial fibrillation ablation; catheter entrapment; phased radiofrequency; pulmonary veins isolation
Mesh:
Year: 2018 PMID: 30019453 DOI: 10.1111/pace.13451