Literature DB >> 30013610

Levels of agreement in the measurements of carotid artery ultrasound across a regional vascular network.

Mervyn Alexander McKenna1, Michelle Clare Bonfield1, Teresa Robinson1.   

Abstract

INTRODUCTION: Duplex ultrasound is often the sole imaging modality used in diagnosing carotid artery disease. However, the reproducibility and repeatability of scientists in determining the peak systolic velocity and end diastolic velocity of the internal carotid artery and common carotid artery (CCA) is widely debated. STUDY AIM: To investigate intra- and inter-operator variability in diagnostic ultrasound of the carotid arteries across a centralised vascular network using a healthy test subject. To identify potential causes of variability and highlight areas for improvement.
METHODS: Fifteen vascular scientists across four hospital Trusts in the Bristol, Bath and Weston vascular network measured the peak systolic velocity and end diastolic velocity of the internal carotid artery and common carotid artery in a subject using a single portable ultrasound machine. A double blind assessment of spectral Doppler images was performed by two vascular clinical scientists for optimal caliper placement, spectral gain and angle correction. Results were compared for intra- and inter-operator variability.
RESULTS: Initial quality assessment of the Doppler images revealed that three out of 15 scientists produced suboptimal results. Box plot analysis of the common carotid artery and internal carotid artery for each scientist revealed significant variance (ANOVA p < 0.05). However, a Levene's test revealed no single operator who consistently produced highly variable results (p = 0.569).
CONCLUSION: This study highlights the difficulty in obtaining consistent velocity measurements from a subject. Despite the variability in absolute peak systolic velocity and end diastolic velocity, scientists were generally consistent in obtaining an optimal spectral Doppler trace. Some issues with consistency were, however, identified which were subsequently addressed.

Entities:  

Keywords:  Diagnostic ultrasound; carotid artery disease; inter-operator variability; vascular network

Year:  2018        PMID: 30013610      PMCID: PMC6042302          DOI: 10.1177/1742271X17751255

Source DB:  PubMed          Journal:  Ultrasound        ISSN: 1742-271X


  16 in total

1.  Finite beam-width ray model for geometric spectral broadening.

Authors:  P R Hoskins; P J Fish; S D Pye; T Anderson
Journal:  Ultrasound Med Biol       Date:  1999-03       Impact factor: 2.998

2.  Formula and nomogram for the sphygmomanometric calculation of the mean arterial pressure.

Authors:  E Meaney; F Alva; R Moguel; A Meaney; J Alva; R Webel
Journal:  Heart       Date:  2000-07       Impact factor: 5.994

3.  Improved interobserver variability and accuracy of echocardiographic visual left ventricular ejection fraction assessment through a self-directed learning program using cardiac magnetic resonance images.

Authors:  Paaladinesh Thavendiranathan; Zoran B Popović; Scott D Flamm; Arun Dahiya; Richard A Grimm; Thomas H Marwick
Journal:  J Am Soc Echocardiogr       Date:  2013-08-28       Impact factor: 5.251

Review 4.  Non-invasive imaging compared with intra-arterial angiography in the diagnosis of symptomatic carotid stenosis: a meta-analysis.

Authors:  J M Wardlaw; F M Chappell; J J K Best; K Wartolowska; E Berry
Journal:  Lancet       Date:  2006-05-06       Impact factor: 79.321

5.  Ultrasound based measurement of 'carotid stenosis >70%': an audit of UK practice.

Authors:  J Walker; A R Naylor
Journal:  Eur J Vasc Endovasc Surg       Date:  2006-01-19       Impact factor: 7.069

6.  Evaluating institutional variability of duplex scanning in the detection of carotid artery stenosis.

Authors:  B K Criswell; M Langsfeld; M J Tullis; J Marek
Journal:  Am J Surg       Date:  1998-12       Impact factor: 2.565

7.  Interobserver and inter-equipment variability of echo-Doppler sonographic evaluation of the superior mesenteric artery.

Authors:  M Zoli; C Merkel; C Sabbà; D Sacerdoti; S Gaiani; G Ferraioli; L Bolondi
Journal:  J Ultrasound Med       Date:  1996-02       Impact factor: 2.153

8.  Variation in duplex peak systolic velocity measurement in a multi-site vascular service.

Authors:  P Normahani; M Aslam; G Martin; N J Standfield; U Jaffer
Journal:  Perfusion       Date:  2015-02-23       Impact factor: 1.972

9.  Benefit of carotid endarterectomy in patients with symptomatic moderate or severe stenosis. North American Symptomatic Carotid Endarterectomy Trial Collaborators.

Authors:  H J Barnett; D W Taylor; M Eliasziw; A J Fox; G G Ferguson; R B Haynes; R N Rankin; G P Clagett; V C Hachinski; D L Sackett; K E Thorpe; H E Meldrum; J D Spence
Journal:  N Engl J Med       Date:  1998-11-12       Impact factor: 91.245

10.  Impact of centralising acute stroke services in English metropolitan areas on mortality and length of hospital stay: difference-in-differences analysis.

Authors:  Stephen Morris; Rachael M Hunter; Angus I G Ramsay; Ruth Boaden; Christopher McKevitt; Catherine Perry; Nanik Pursani; Anthony G Rudd; Lee H Schwamm; Simon J Turner; Pippa J Tyrrell; Charles D A Wolfe; Naomi J Fulop
Journal:  BMJ       Date:  2014-08-05
View more
  1 in total

Review 1.  Why are we still debating criteria for carotid artery stenosis?

Authors:  Victor J Del Brutto; Heather L Gornik; Tatjana Rundek
Journal:  Ann Transl Med       Date:  2020-10
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.