Takayuki Tamura1, Miyuki Takasu2, Toru Higaki2,3, Kazushi Yokomachi1, Yuji Akiyama1, Hiroomi Sumida1, Yasushi Nagata1,3, Kazuo Awai2,3.
Abstract
PURPOSE: The aim of this study was to compare the tumor conspicuity on actual measured diffusion-weighted images (aDWIs) and computed DWI (cDWI) of human breast tumors and to examine, by use of a phantom, whether cDWI improves their conspicuity.Entities:
Keywords: breast tumor; computed diffusion-weighted image; multi-exponential diffusion-weighted image decay
Mesh:
Year: 2018 PMID: 30012905 PMCID: PMC6460120 DOI: 10.2463/mrms.mp.2018-0011
Source DB: PubMed Journal: Magn Reson Med Sci ISSN: 1347-3182 Impact factor: 2.471
Fig. 1.Actual measured diffusion-weighted image (aDWI) and computed DWI (cDWI) of a 76-year-old woman with invasive ductal carcinoma. (a) Upper: aDWI, lower: cDWI (b = 0 and 700 s/mm2). (b) 3D T1-weighted fast gradient-echo image obtained after the administration of contrast medium. (c and d) ROI position (c) and signal decay of aDWI and cDWI (d).
Fig. 2.Schema of our phantom mimicking a breast tumor and normal mammary gland. The bi-exponential decay on diffusion-weighted images (DWIs) was constructed by scanning the interface between the fast and the slow component using the partial volume effect. ADC, apparent diffusion coefficient.
Fig. 3.Comparison of the signal-to-noise ratio (SNR) on computed diffusion-weighted images (cDWIs) and actual measured DWIs (aDWIs) obtained at different b-values. The SNR rate was calculated by dividing the SNR on cDWIs by the SNR on aDWIs.
Fig. 4.Phantom study – diffusion-weighted image (DWI). (a) Actual measured multiple b-value DWI series. (b) High b-value cDWI (b = 0 and 1000 s/mm2). (c–e) Single b-value series. TE is shortened and the number of excitations (NEX) is 2, 4, and 8.
Fig. 5.Graph showing the signal decay of the fast (a) and the slow component (b) on actual measured diffusion-weighted image (aDWI). The tumor phantom is presented in (c) and the background in (d).
Fig. 6.Comparison of the signal-to-noise ratio (SNR) of the background (a) and the tumor phantom (b) and comparison of the contrast (c) and the contrast-to-noise ratio (CNR) (d) on images of the tumor phantom and the background. aDWI, actual measured diffusion-weighted image; cDWI, computed diffusion-weighted image.