Literature DB >> 29998648

Investigating optical path and differential pathlength factor in reflectance photoplethysmography for the assessment of perfusion.

Subhasri Chatterjee1, Tomas Ysehak Abay1, Justin Paul Phillips1, Panayiotis A Kyriacou1.   

Abstract

Photoplethysmography (PPG) is an optical noninvasive technique with the potential for assessing tissue perfusion. The relative time-change in the concentration of oxyhemoglobin and deoxyhemoglobin in the blood can be derived from DC part of the PPG signal. However, the absolute concentration cannot be determined due to the inadequate data on PPG optical paths. The optical path and differential pathlength factor (DPF) for PPG at red (660 nm) and infrared (880 nm) wavelengths were investigated using a heterogeneous Monte Carlo model of the human forearm. Using the simulated DPFs, the absolute time-change in concentrations were determined from PPG signals recorded from the same tissue site. Results were compared with three conditions of approximated DPFs. Results showed the variation of the optical-path and DPF with different wavelengths and source-detector separations. Approximations resulted in significant errors, for example, using NIRS DPF in PPG led to "cross talk" of -0.4297 and 0.060 and an error of 15.16% to 25.18%. Results confirmed the feasibility of using the PPG (DC) for the assessment of tissue perfusion. The study also identified the inappropriateness of the assumption that DPF is independent of wavelength or source-detector separations and set the platform for further studies on investigating optical pathlengths and DPF in PPG. (2018) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE).

Entities:  

Keywords:  Beer–Lambert law; Monte Carlo model; near-infrared spectroscopy; optical path; oxy/deoxy hemoglobin; photoplethysmography

Year:  2018        PMID: 29998648     DOI: 10.1117/1.JBO.23.7.075005

Source DB:  PubMed          Journal:  J Biomed Opt        ISSN: 1083-3668            Impact factor:   3.170


  5 in total

1.  Assessing pulse transit time to the skeletal muscle microcirculation using near-infrared spectroscopy.

Authors:  Cody P Anderson; Song-Young Park
Journal:  J Appl Physiol (1985)       Date:  2022-07-14

2.  Development and validation of quantitative optical index of skin blood content.

Authors:  Yu-Hao Peng; Jean-Michel I Maarek
Journal:  J Biomed Opt       Date:  2022-06       Impact factor: 3.758

3.  A Computational Modeling and Simulation Workflow to Investigate the Impact of Patient-Specific and Device Factors on Hemodynamic Measurements from Non-Invasive Photoplethysmography.

Authors:  Jesse Fine; Michael J McShane; Gerard L Coté; Christopher G Scully
Journal:  Biosensors (Basel)       Date:  2022-08-04

4.  Noninvasive Portable Hemoglobin Concentration Monitoring System Using Optical Sensor for Anemia Disease.

Authors:  Chuchart Pintavirooj; Baorong Ni; Chaiwat Chatkobkool; Kittitorn Pinijkij
Journal:  Healthcare (Basel)       Date:  2021-05-29

5.  Novel Polydimethylsiloxane (PDMS) Pulsatile Vascular Tissue Phantoms for the In-Vitro Investigation of Light Tissue Interaction in Photoplethysmography.

Authors:  Michelle Nomoni; James M May; Panayiotis A Kyriacou
Journal:  Sensors (Basel)       Date:  2020-07-30       Impact factor: 3.576

  5 in total