CMOS High-Voltage Analog 1-64 Multiplexer/Demultiplexer for Integrated Ultrasound Guided Breast Needle Biopsy.

Hayong Jung, Robert Wodnicki, Hae Gyun Lim, Chi Woo Yoon, Bong Jin Kang, Changhan Yoon, Changyang Lee, Jae Youn Hwang, Hyung Ham Kim, Hojong Choi, Mike Shuo-Wei Chen, Qifa Zhou, K Kirk Shung.   

Abstract

Ultrasound guided needle biopsy is an important method for collection of breast cancer tissue. In this paper, we report on the design and testing of a high-voltage 1 to 64 Multiplexer/Demultiplexer (MUX/De-MUX) integrated circuit (IC) for ultrasound-guided breast biopsy applications implemented in a high-voltage CMOS process. The IC is intended to be incorporated inside the breast biopsy needle and is designed to fit inside the needle inner diameter of 2.38 mm. The MUX/De-MUX electronics are made up of three parts, including a low-voltage 6 to 64 decoder, a level shifter to convert from low voltage to high voltage, and analog high-voltage switches. Experimental results show a -3-dB bandwidth of over 70 MHz, Rds (on) of , -2.279-dB insertion loss, and -17.5-dB off isolation at 70 MHz with low-voltage input. Finally, we present results obtained via synthetic aperture imaging using the fabricated MUX/De-Mux device and a high-frequency ultrasound array. This device and technique hold promise for high-frequency imaging probes where a limited number of elements are used and the depth of penetration is short such as in breast biopsy and intravascular applications.

Entities:  

Mesh:

Year:  2018        PMID: 29994523     DOI: 10.1109/TUFFC.2018.2837127

Source DB:  PubMed          Journal:  IEEE Trans Ultrason Ferroelectr Freq Control        ISSN: 0885-3010            Impact factor:   2.725


  3 in total

Review 1.  Current Ultrasound Technologies and Instrumentation in the Assessment and Monitoring of COVID-19 Positive Patients.

Authors:  Xuejun Qian; Robert Wodnicki; Haochen Kang; Junhang Zhang; Hisham Tchelepi; Qifa Zhou
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2020-08-28       Impact factor: 2.725

2.  A Robotically Steerable Guidewire With Forward-Viewing Ultrasound: Development of Technology for Minimally-Invasive Imaging.

Authors:  Graham C Collins; Achraj Sarma; Zachary L Bercu; Jaydev P Desai; Brooks D Lindsey
Journal:  IEEE Trans Biomed Eng       Date:  2021-06-17       Impact factor: 4.756

3.  Highly Integrated Multiplexing and Buffering Electronics for Large Aperture Ultrasonic Arrays.

Authors:  Robert Wodnicki; Haochen Kang; Di Li; Douglas N Stephens; Hayong Jung; Yizhe Sun; Ruimin Chen; Lai-Ming Jiang; Nestor E Cabrera-Munoz; Josquin Foiret; Qifa Zhou; Katherine W Ferrara
Journal:  BME Front       Date:  2022-06-30