An Algorithm for Heart Rate Extraction From Acoustic Recordings at the Neck.

Piyush Sharma, Syed Anas Imtiaz, Esther Rodriguez-Villegas.   

Abstract

Heart rate is an important physiological parameter to assess the cardiac condition of an individual and is traditionally determined by attaching multiple electrodes on the chest of a subject to record the electrical activity of the heart. The installation and handling complexities of such systems does not prove feasible for a user to undergo a long-term monitoring in the home settings. A small-sized, battery-operated wearable monitoring device is placed on the suprasternal notch at neck to record acoustic signals containing information about breathing and cardiac sounds. The heart sounds obtained are heavily corrupted by the respiratory cycles and other external artifacts. This paper presents a novel algorithm for reliably extracting the heart rate from such acoustic recordings, keeping in mind the constraints posed by the wearable technology. The methodology constructs the Hilbert energy envelope of the signal by calculating its instantaneous characteristics to segment and classify a cardiac cycle into S1 and S2 sounds using their timing characteristics. The algorithm is tested on a dataset consisting of 13 subjects with an approximate data length of 75 h and achieves an accuracy of 94.34%, an RMS error of 3.96 bpm and a correlation coefficient of 0.93 with reference to a commercial device in use.

Year:  2018        PMID: 29993496     DOI: 10.1109/TBME.2018.2836187

Source DB:  PubMed          Journal:  IEEE Trans Biomed Eng        ISSN: 0018-9294            Impact factor:   4.538


  3 in total

1.  Tracheal Sound Analysis.

Authors:  AbdelKebir Sabil; Sandrine Launois
Journal:  Adv Exp Med Biol       Date:  2022       Impact factor: 3.650

2.  Construction of Physical Education Quality Evaluation Index and Analysis with Wearable Device.

Authors:  Lei Fang
Journal:  Comput Intell Neurosci       Date:  2022-05-26

3.  Acoustic Sensing as a Novel Wearable Approach for Cardiac Monitoring at the Wrist.

Authors:  Piyush Sharma; Syed Anas Imtiaz; Esther Rodriguez-Villegas
Journal:  Sci Rep       Date:  2019-12-27       Impact factor: 4.379