Literature DB >> 29987949

Establishing thresholds for metal ion levels in patients with bilateral Articular Surface Replacement hip arthroplasty.

Gabrielle S Donahue1, Vincent P Galea1, Inari Laaksonen1, James W Connelly1, Orhun Muratoglu1, Henrik Malchau1.   

Abstract

INTRODUCTION: The primary aim of the current study was to determine whether metal ion levels could identify well- from poorly-functioning implants in patients after undergoing bilateral Articular Surface Replacement (ASR) Hip System. The secondary aim was to establish threshold values for Co and Cr that could be used to predict clinical problems in these patients.
MATERIALS AND METHODS: The study included 125 patients treated with bilateral ASR hip resurfacing (HRA) and 93 patients treated with bilateral ASR XL total hip arthroplasty (THA) at a mean follow-up of 7.1 years post-surgery. Patients were labelled as well- or poorly-functioning based on strict criteria. We calculated the Co and Cr safe upper ion limits (SUL) for ASR HRA and ASR XL THA separately and assessed these cutoffs ability to predict clinical problems.
RESULTS: In patients with bilateral ASR HRA, the well-functioning group had lower ions than the poorly-functioning. In ASR THA patients, there was no difference in blood Co or Cr levels between well- and poorly-functioning implant groups (p = 0.118 and p = 0.379, respectively). Both Co and Cr could effectively discriminate between well-functioning and poorly-functioning ASR HRA implants (p < 0.001 and p < 0.001, respectively), but not for ASR THA implants (p = 0.118 and p = 0.564, respectively). SULs of 4.0 ppb for Co and 4.0 ppb for Cr could differentiate between well- and poorly-functioning implants with high specificity.
CONCLUSIONS: Metal ion levels are a more reliable tool in MoM HRA than THA when discriminating well- from poorly-functioning implants in bilateral patients. For patients treated with bilateral ASR HRA, ion levels above 4.0 ppb are highly specific for the detection of clinical problems.

Entities:  

Keywords:  Articular Surface Replacement; blood metal ion levels; hip resurfacing; metal-on-metal; total hip arthroplasty

Mesh:

Substances:

Year:  2018        PMID: 29987949     DOI: 10.1177/1120700018784273

Source DB:  PubMed          Journal:  Hip Int        ISSN: 1120-7000            Impact factor:   2.135


  2 in total

Review 1.  The contribution of the histopathological examination to the diagnosis of adverse local tissue reactions in arthroplasty.

Authors:  Giorgio Perino; Ivan De Martino; Lingxin Zhang; Zhidao Xia; Jiri Gallo; Shonali Natu; David Langton; Monika Huber; Anastasia Rakow; Janosch Schoon; Enrique Gomez-Barrena; Veit Krenn
Journal:  EFORT Open Rev       Date:  2021-06-28

2.  Repeated cobalt and chromium ion measurements in patients with bilateral large-diameter head metal-on-metal ReCap-M2A-Magnum total hip replacement.

Authors:  Sakari Pietiläinen; Heikki Mäntymäki; Tero Vahlberg; Aleksi Reito; Antti Eskelinen; Petteri Lankinen; Keijo Mäkelä
Journal:  Acta Orthop       Date:  2020-04-14       Impact factor: 3.717