Incorporating a companion test into the noninferiority design of medical device trials.

Nelson T Lu1, Yunling Xu1, Ying Yang1.   

Abstract

Noninferiority trials are commonly utilized to evaluate the safety and effectiveness of medical devices. It could happen that the noninferiority hypothesis is rejected while the performance of the active control is clinically not satisfactory. This may pose a great challenge when making a regulatory decision. To avoid such a difficult situation, we propose to conduct a companion test to assess the performance of the active control when testing the main noninferiority hypothesis and to incorporate such a test into the study design. Under our proposal, the noninferiority of the investigational device to the active control can only be claimed when both hypotheses are rejected. The operating characteristics of the proposed study design based on these two tests can be fully evaluated at the design stage. This proposed approach is aimed to facilitate regulatory decision making in a more transparent manner.

Keywords:  Active control performance; companion test; medical device; noninferiority; transparency

Mesh:

Year:  2018        PMID: 29985744     DOI: 10.1080/10543406.2018.1489403

Source DB:  PubMed          Journal:  J Biopharm Stat        ISSN: 1054-3406            Impact factor:   1.051