| Literature DB >> 29985149 |
Alessandro Mengarelli, Stefano Cardarelli, Annachiara Strazza, Francesco Di Nardo, Sandro Fioretti, Federica Verdini.
Abstract
The functional reach test (FRT) is widely used for assessing dynamic balance stability in elderly and pathological subjects. Force platforms (FPs) represent a fundamental part of the instrumented FRT experimental setup due to the central role of center-of-pressure (COP) displacement in FRT analysis. Recently, the nintendo wii balance board (NBB) has been suggested as a low-cost and reliable device for ground reaction force and COP measurement in poorly dynamic motor tasks. Therefore, this paper aimed to compare NBB-COP data with those obtained from a laboratory-grade platform during FRT. Data from 48 healthy subjects were simultaneously acquired from both devices. FP-COP and NBB-COP trajectories showed a remarkable correlation in both directions ( ) and low root-mean-square error values (1.14 ± 0.88 mm and 0.55 ± 0.28 mm for anterior-posterior and medial-lateral direction). Fixed biases between COP-based parameters did not exceed 2% of the FP outcomes with high consistency throughout the present measurement range (ICC consistency always >0.950). Only the COP mean velocity exhibited a tendency toward proportional errors, which can be adjusted by a calibration of NBB data. Findings of this paper confirmed the NBB validity for COP measurement in a widely used motor task as the functional reach, supporting the feasibility of NBB in research scenarios.Mesh:
Year: 2018 PMID: 29985149 DOI: 10.1109/TNSRE.2018.2843884
Source DB: PubMed Journal: IEEE Trans Neural Syst Rehabil Eng ISSN: 1534-4320 Impact factor: 3.802