Andrea L Hergenroeder1, Bethany Barone Gibbs1, Mary P Kotlarczyk1, Robert J Kowalsky2, Subashan Perera1, Jennifer S Brach1.
Abstract
Objective: The aim of this study is to evaluate accuracy of research activity monitors in measuring steps in older adults with a range of walking abilities. Method: Participants completed an initial assessment of gait speed. The accuracy of each monitor to record 100 steps was assessed across two walking trials.Entities:
Keywords: accuracy; activity monitor; older adults; step count
Year: 2018 PMID: 29977979 PMCID: PMC6024488 DOI: 10.1177/2333721418781126
Source DB: PubMed Journal: Gerontol Geriatr Med ISSN: 2333-7214
Characteristics of the Activity Monitors.
| Activity monitor | Measures | Wear location | Mechanism |
|---|---|---|---|
| Actigraph GT3X | Activity intensity, energy expenditure, activity count, step count | Waist | Accelerometer |
| ActivPAL | Sedentary minutes, standing minutes, stepping minutes, transitions, energy expenditure, step count | Thigh | Accelerometer inclinometer |
| StepWatch | Step count | Ankle | Accelerometer |
Participant Characteristics (n = 43).
| Characteristic | |
|---|---|
| Age, years | 86.7 ± 5.7 (69-98) |
| Female | 35 (81.4%) |
| Race | |
| White | 43 (100.0%) |
| Education | |
| Elementary | 1 (2.3%) |
| High school | 23 (53.5%) |
| College | 16 (37.2%) |
| Postgraduate | 3 (7.0%) |
| Comorbid conditions | |
| Number of comorbidities | 3.0 ± 1.2 (1-6) |
| Comorbidity domains | |
| Cardiovascular | 7 (16.3%) |
| Neurological | 7 (16.3%) |
| Musculoskeletal | 36 (83.7%) |
| General | 15 (34.9%) |
| Visual/hearing | 42 (97.7%) |
| Diabetes | 8 (18.6%) |
| Cancer | 10 (23.3%) |
| Lung | 5 (11.6%) |
| Height (m) | 1.61 ± 0.09 |
| Weight (kg) | 68.3 ± 13.4 |
| Body mass index (kg/m2) | 26.1 ± 4.1 |
| Gait speed (m/s) | 0.84 ± 0.24 |
| Assistive device | |
| No device | 22 (51.2%) |
| Cane | 4 (9.3%) |
| Wheeled walker/rollator | 17 (39.5%) |
Figure 1.Overall percentage accuracy of activity monitors.
Percentage Accuracy of Activity Monitors Stratified by Assistive Device Use.
| Activity monitor | No device | Wheeled Walker/rollator | Cane | |
|---|---|---|---|---|
| Actigraph GT3X | 67.3 (35.6) | 32.9 (28.8) | 43.0 (26.4) | .0071 |
| ActivPAL | 96.3 (7.0) | 92.7 (10.2) | 80.6 (27.3) | .0392 |
| StepWatch | 99.1 (1.0) | 99.2 (0.7) | 97.7 (4.4) | .7312 |
Note. Accuracy defined as monitor steps/observed steps × 100 with 100% indicating perfect accuracy.
Percentage Accuracy of Activity Monitors Stratified by Gait Speed Category.
| Activity monitor | <0.6 m/s | 0.60-0.79 m/s | 0.8-1.0 m/s | >1.0 m/s | |
|---|---|---|---|---|---|
| Actigraph GT3X | 14.1 (18.6) | 35.6 (31.2) | 52.7 (28.6) | 85.1 (20.5) | <.0001 |
| ActivPAL | 86.8 (14.0) | 91.8 (15.1) | 98.2 (0.8) | 95.1 (9.2) | .0174 |
| StepWatch | 99.4 (0.6) | 98.2 (2.5) | 99.7 (0.3) | 99.0 (0.9) | .0208 |
Note. Accuracy defined as monitor steps/observed steps × 100 with 100% indicating perfect accuracy.