| Literature DB >> 29963459 |
Jorge Aguiar1, José A Silva2, Gerardo García3, Gerardo Guillén1, Julio C Aguilar1.
Abstract
AIM: This research focused on the results of the cross-validation program related with the performance of a Cuban novel low-cost real-time quantitative polymerase chain reaction (qPCR) assay for hepatitis B virus (HBV) quantification developed by the Therapeutic Vaccine against Hepatitis B Department, Vaccines Division, Center for Genetic Engineering and Biotechnology (CIGB), Havana, Cuba.Entities:
Keywords: Crossing point; Cycles threshold; Hepatitis B virus; Laboratory research; Plasmid; Primer; Quantitative polymerase chain reaction.
Year: 2018 PMID: 29963459 PMCID: PMC6024048 DOI: 10.5005/jp-journals-10018-1255
Source DB: PubMed Journal: Euroasian J Hepatogastroenterol ISSN: 2231-5047
Table 1: Mean of Ct values for each PCR instrument
| 900,000 c/r | 15.40 | 15.55 | 16.90 | 15.81 | 14.49 | 15.63 | 0.87 | 5.54 | |||||||||
| 90,000 c/r | 18.94 | 18.28 | 20.51 | 18.59 | 17.34 | 18.73 | 1.16 | 6.18 | |||||||||
| 9,000 c/r | 22.71 | 21.93 | 24.27 | 22.77 | 20.84 | 22.50 | 1.26 | 5.59 | |||||||||
| 900 c/r | 26.51 | 25.2 | 27.73 | 26.45 | 24.18 | 26.01 | 1.36 | 5.23 | |||||||||
| 90 c/r | 29.55 | 27.81 | 31.25 | 30.03 | 26.95 | 29.12 | 1.73 | 5.94 | |||||||||
| 9 c/r | 33.00 | 29.62 | 35.37 | 32.72 | 28.40 | 31.82 | 2.80 | 8.80 | |||||||||
| 0.9 c/r | 34.82 | 29.96 | 36.86 | 35.38 | 29.18 | 33.24 | 3.44 | 10.36 | |||||||||
| 0.09 c/r | 39.10 | 30.96 | 38.73 | 42.33 | 29.70 | 36.16 | 5.52 | 15.28 |
SD: Standard deviation
Table 2: Polymerase chain reaction efficiency for each PCR instrument
| Efficiency (%) | 99.9 | 102.4 | 95.0 | 96.5 | 106.4 | ||||||
| R[ | 0.984 | 0.978 | 0.990 | 0.995 | 0.997 | ||||||
| Slope | –3.323 | –3.265 | –3.449 | –3.409 | –3.177 | ||||||
| Y-intercept | 35.66 | 34.79 | 37.75 | 35.67 | 27.22 | ||||||
| Dynamic | Until | Until | Until | Until 0.9 | Until | ||||||
| range | 0.09 c/r | 9 c/r | 0.9 c/r | c/r | 90 c/r |
R2 Squared correlation or linear regression. Note that the dynamic range varies among the PCR instruments
Table 3: Reproducibility of HBV assay
| 1 | 13.25 | 29.63 | 12.85 | 29.90 | 14.54 | 31.60 | 13.05 | 29.88 | 12.87 | 29.62 | |||||||||||
| 2 | 13.22 | 29.44 | 12.74 | 30.24 | 14.49 | 31.81 | 12.79 | 30.02 | 12.77 | 29.68 | |||||||||||
| 3 | 13.16 | 29.97 | 12.73 | 29.98 | 14.23 | 31.87 | 13.11 | 30.45 | 12.75 | 29.45 | |||||||||||
| 4 | 13.08 | 30.05 | 12.74 | 29.67 | 14.29 | 31.77 | 12.96 | 30.11 | 13.00 | 29.40 | |||||||||||
| 5 | 13.00 | 30.00 | 12.76 | 29.84 | 14.33 | 31.58 | 13.12 | 30.21 | 12.84 | 29.77 | |||||||||||
| 6 | 12.99 | 30.03 | 12.85 | 29.51 | 14.79 | 31.56 | 42.12 | 29.47 | 12.82 | 30.11 | |||||||||||
| 7 | 13.41 | 29.94 | 12.90 | 29.83 | 14.84 | 31.35 | 12.88 | 30.38 | 13.22 | 29.54 | |||||||||||
| 8 | 13.30 | 30.36 | 12.72 | 29.75 | 15.09 | 32.38 | 13.14 | 30.20 | 12.98 | 29.48 | |||||||||||
| 9 | 13.13 | 29.89 | 12.65 | 29.45 | 14.64 | 32.36 | 13.61 | 29.62 | 12.82 | 29.38 | |||||||||||
| 10 | 13.25 | 30.28 | 12.64 | 29.52 | 14.75 | 31.75 | 13.13 | 30.19 | 12.75 | 29.67 | |||||||||||
| Mean Ct | 13.18 | 29.96 | 12.76 | 29.77 | 14.60 | 31.80 | 13.09 | 30.05 | 12.88 | 29.61 | |||||||||||
| SD | 0.13 | 0.27 | 0.09 | 0.24 | 0.27 | 0.33 | 0.23 | 0.31 | 0.15 | 0.22 | |||||||||||
| CV (%) | 1.01 | 0.90 | 0.67 | 0.82 | 1.88 | 1.05 | 1.77 | 1.05 | 1.14 | 0.74 | |||||||||||
| M6 CV cycler (%) | 1.29 | ||||||||||||||||||||
| M19 CV cycler (%) | 0.91 | ||||||||||||||||||||
Cycles threshold values from 10 replicates were evaluated for each sample and instrument; SD: Standard deviation; CV (%): Coefficient of variation expressed in %
Table 4: Performance of the different primer suppliers (CIGB and BCD) on the Roche LightCycler® 480 II
| 900,000 c/r | 15.55 | 15.59 | 0.05 | 0.07 | 0.34 | 0.42 | |||||||
| 90,000 c/r | 18.28 | 18.43 | 0.05 | 0.14 | 0.28 | 0.75 | |||||||
| 9,000 c/r | 21.93 | 21.89 | 0.06 | 0.03 | 0.28 | 0.14 | |||||||
| 900 c/r | 25.2 | 25.14 | 0.12 | 0.14 | 0.46 | 0.55 | |||||||
| 90 c/r | 27.81 | 27.94 | 0.08 | 0.14 | 0.29 | 0.51 | |||||||
| 9 c/r | 29.62 | 29.51 | 0.12 | 0.27 | 0.39 | 0.92 | |||||||
| 0.9 c/r | 29.96 | 29.96 | 0.20 | 0.04 | 0.65 | 0.13 | |||||||
| 0.09 c/r | 30.96 | 30.55 | 0.49 | 0.04 | 1.58 | 0.14 | |||||||
Dilution series with the pST012012 plasmid at concentrations of 900,000 to 0.09 c/r were used, and three replicates per dilution step were measured in the qPCR. The means of Ct or Cp values were compared between the thermocyclers. SD: Standard deviation; CV (%): Coefficient of variation expressed in percent
Table 5: Performance of the different primer suppliers (CIGB and BCD) on the Roche LightCycler® 480 II
| Efficiency (%) | 102.4 | 108.1 | |||
| R[ | 0.978 | 0.986 | |||
| Slope | –3.265 | –3.142 | |||
| Y-intercept | 34.79 | 34.23 | |||
| Dynamic range | Until 9 c/r | Until 9 c/r |
PCR: Polymerase chain reaction efficiency expressed in percent was compared between the primer suppliers; R[2]: Squared correlation