Literature DB >> 29962269

Balloon-Based Organ Retractor With Increased Safety and Reduced Invasiveness During Video-Assisted Thoracoscopic Surgery.

Daisuke Taniguchi1,2, Keitaro Matsumoto1,2, Yoshihiro Kondo2,3, Koichiro Shimoyama1,2, Yasumasa Hashimoto1,2, Abdelmotagaly Elgalad1,2, Tomohiro Obata1,2, Takuro Miyazaki1, Tomoshi Tsuchiya1, Naoto Matsuo1,2, Murray John Lawn1,2, Naoya Yamasaki1,2, Ikuo Yamamoto2,3, Takeshi Nagayasu1,2.   

Abstract

OBJECTIVES: In recent years, video-assisted thoracoscopic surgery (VATS) has increasingly become the preferred technique for thoracic surgery. However, the inherent characteristics of the lungs as large, soft, slippery, and delicate creates difficulties for pulmonary surgery. In this article, we outline the development and assessment of a balloon-based organ retractor for VATS via collaboration between medical and engineering personnel.
METHODS: A dry lab trial and accompanying questionnaire assessment were performed by a group of thoracic surgeons. Objective pressure measurements were obtained, and animal experiment on pigs was performed.
RESULTS: In the dry lab trial, use of the developed organ retractor required significantly less time and resulted in fewer difficulties than using a Cherry Dissector. The measured pressure per mm2 of the developed retractor was clearly lower than that for the Cherry Dissector. The questionnaire completed by the surgeons following the dry lab and animal experiments showed that most of the surgeons (7 surgeons out of 9) were satisfied with the quality of the balloon-based retractor based on a score of 3.13 ± 0.28 (mean ± standard deviation) out of 4.0. During the animal experiment, the balloon-based retractor provided stable and clear viewing with minimal need for adjustment.
CONCLUSION: This balloon-based retractor could contribute to increased safety and less-invasive VATS.

Entities:  

Keywords:  VATS; inflatable balloon mechanism; medical-engineering collaboration; organ retractor; video-assisted thoracoscopic surgery

Mesh:

Year:  2018        PMID: 29962269     DOI: 10.1177/1553350618784473

Source DB:  PubMed          Journal:  Surg Innov        ISSN: 1553-3506            Impact factor:   2.058


  2 in total

1.  The usefulness and utilization of a detachable steel wire-rimmed retractor (KN retractor) for endoscopic thyroid and parathyroid surgery.

Authors:  Akihiro Nakajo; Koji Minami; Yoshiaki Shinden; Tadahiro Hirashima; Hazuki Saho; Yuki Nomoto; Ayako Nagata; Takao Ohtsuka
Journal:  Surg Today       Date:  2020-09-02       Impact factor: 2.549

2.  Single-incision laparoscopic total colectomy using an organ retractor: A case report.

Authors:  Kazuhide Ishimaru; Tetsuro Tominaga; Takashi Nonaka; Makoto Hisanaga; Akiko Fukuda; Takafumi Yoshimoto; Daiki Takei; Shigekazu Hidaka; Terumitsu Sawai; Takeshi Nagayasu
Journal:  Ann Med Surg (Lond)       Date:  2020-05-20
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.